9610617-2007-00029

MDR report key: 981403 · Received November 21, 2007

Report

Report Number
9610617-2007-00029
Event Type
Other
Date Received
November 21, 2007
Product Code
FFK
Report Source
Manufacturer report

Narratives

Additional Manufacturer Narrative · 1

CHECKED ALL FUNCTIONS THE ONLY PROBLEM FOUND WITH THE INSTRUMENT 27076A WAS THE LEVER ON THE LOCKING COLLAR FOR THE TELESCOPE IS STRIPPED AND NEEDS REPLACING. THE LOCKING LEVER HAS NO EFFECT ON THE ASSEMBLY OR ALIGNMENT OF THE JAWS ON THIS INSTRUMENT; THE ONLY FUNCTION IT HAS IS TO LOCK THE SCOPE IN THE CHANNEL ONCE THE INSTRUMENT IS ASSEMBLED. THERE WAS NO PROBLEM FOUND WITH THE FUNCTIONING OF THE INSTRUMENT. HOSPITAL ALSO REPORTED THAT THEY BELIEVE THERE WAS NO MALFUNCTION OF THE INSTRUMENT, BUT RATHER A POSSIBLE USER ERROR IN NOT TIGHTENING FRONT SCREW PEG WHICH WOULD HAVE KEPT JAWS ALIGNED CORRECTLY AND MADE REMOVAL POSSIBLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FFK

Patients

Seq Age Sex Outcome Treatment
1