FDA Adverse Event Death Summary report: N

BJORK-SHILEY 60 DEGREE CONVEXO-CONCAVE

MDR report key: 981367 · Received January 15, 2008

Report

Report Number
6000012-2007-00013
Event Type
Death
Date Received
January 15, 2008
Date of Event
November 27, 2007
Report Date
December 19, 2007
Manufacturer
SHILEY, INC.
Product Code
LWQ
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE VALVE WAS EXAMINED WITH A LIGHT MICROSCOPE AT 20X MAGNIFICATION. THE RIGHT LEG OF THE OUTLET STRUT WAS FOUND TO HAVE SEPARATED FROM THE VALVE FLANGE (TERMED A "SINGLE LEG SEPARATION"). THE LEFT LEG OF THE OUTLET STRUT WAS FOUND TO BE INTACT. ACCORDING TO THE VALVE EXAMINATION REPORT, WEAR PATTERNS, SCRATCHES, AND INSTRUMENT MARKS TYPICAL FOR VALVES IMPLANTED FOR PERIODS GREATER THAN 25 YEARS WERE NOTED ON THE OUTLET STRUT, INLET STRUT, FLANGE INNER DIAMETER, RIMS, AND DISC.

Description of Event or Problem · 1

THE PT CONTACTED SHILEY TO REPORT THAT HIS BJORK-SHILEY CONVEXO-CONCAVE AORTIC AND MITRAL VALVES WOULD BE ELECTIVELY REPLACED AND REQUESTED TO HAVE THE VALVES RETURNED FOR EXAMINATION. MICROSCOPIC EXAMINATION REVEALED A SINGLE LEG SEPARATION OF THE RIGHT OUTLET STRUT LEG OF THE MITRAL VALVE. SUBSEQUENTLY, SHILEY WAS NOTIFIED THAT THE PT DIED 7 DAYS FOLLOWING VALVE REPLACEMENT SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BJORK-SHILEY 60 DEGREE CONVEXO-CONCAVE HEART VALVE LWQ SHILEY, INC. 29MBRC NA

Patients

Seq Age Sex Outcome Treatment
1 50 YR Death| H (STOP'D)| BJORK-SHILEY 60 DEGREE CONVEXO-CONCAVE AORTIC