BJORK-SHILEY 60 DEGREE CONVEXO-CONCAVE
Report
- Report Number
- 6000012-2007-00013
- Event Type
- Death
- Date Received
- January 15, 2008
- Date of Event
- November 27, 2007
- Report Date
- December 19, 2007
- Manufacturer
- SHILEY, INC.
- Product Code
- LWQ
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THE VALVE WAS EXAMINED WITH A LIGHT MICROSCOPE AT 20X MAGNIFICATION. THE RIGHT LEG OF THE OUTLET STRUT WAS FOUND TO HAVE SEPARATED FROM THE VALVE FLANGE (TERMED A "SINGLE LEG SEPARATION"). THE LEFT LEG OF THE OUTLET STRUT WAS FOUND TO BE INTACT. ACCORDING TO THE VALVE EXAMINATION REPORT, WEAR PATTERNS, SCRATCHES, AND INSTRUMENT MARKS TYPICAL FOR VALVES IMPLANTED FOR PERIODS GREATER THAN 25 YEARS WERE NOTED ON THE OUTLET STRUT, INLET STRUT, FLANGE INNER DIAMETER, RIMS, AND DISC.
THE PT CONTACTED SHILEY TO REPORT THAT HIS BJORK-SHILEY CONVEXO-CONCAVE AORTIC AND MITRAL VALVES WOULD BE ELECTIVELY REPLACED AND REQUESTED TO HAVE THE VALVES RETURNED FOR EXAMINATION. MICROSCOPIC EXAMINATION REVEALED A SINGLE LEG SEPARATION OF THE RIGHT OUTLET STRUT LEG OF THE MITRAL VALVE. SUBSEQUENTLY, SHILEY WAS NOTIFIED THAT THE PT DIED 7 DAYS FOLLOWING VALVE REPLACEMENT SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BJORK-SHILEY 60 DEGREE CONVEXO-CONCAVE | HEART VALVE | LWQ | SHILEY, INC. | 29MBRC | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | Death| H | (STOP'D)| BJORK-SHILEY 60 DEGREE CONVEXO-CONCAVE AORTIC |