- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FOLLOWING FIELDS HAVE BEEN UPDATED WITH CORRECTIONS: H.3. REASON CODE FOR NO EVALUATION: OTHER. H.3. IF OTHER SPECIFY: SEE H.10 H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT BD ULTRA FINE¿ PEN NEEDLES HAD NO INSULIN FLOW DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 320122 BATCH NO. 9162522 IT WAS REPORTED THAT THERE WAS NO INSULIN FLOW DURING INJECTION. VERBATIM: CONSUMER STATED, NO INSULIN FLOW WHEN TAKING INJECTION. STATED, SHE DOES NOT PRIME BEFORE INJECTION. STATED, SHE DOES NOT WANT TO WASTE INSULIN PRIMING.
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. INVESTIGATION SUMMARY: LEVEL B INVESTIGATION - COMPLAINT EVALUATION / COMPLAINT HISTORY CHECK FOR THE EVENT(S) THAT OCCURRED. SEVERITY: S_2__; OCCURRENCE: A COMPLAINT HISTORY CHECK WAS PERFORMED AND THIS IS THE 4TH RELATED COMPLAINT FOR NEEDLE CLOG ON LOT # 9162522. INVESTIGATION SUMMARY: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. IF SAMPLES ARE RECEIVED IN THE FUTURE THE COMPLAINT WILL BE REOPENED FOR FURTHER INVESTIGATION. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCE'S WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. INVESTIGATION CONCLUSION: AS NO SAMPLES AND/OR PHOTO(S) WERE RECEIVED THE INVESTIGATION CONCLUDED: UNCONFIRMED: BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE AS NO SAMPLES OR PHOTOS WERE RETURNED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. ROOT CAUSE DESCRIPTION: ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED AS NO SAMPLES OR PHOTOS WERE RETURNED. RATIONALE: BASED ON THE INVESTIGATION, NO ADDITIONAL INVESTIGATION AND NO CAPA IS REQUIRED AT THIS TIME.
IT WAS REPORTED THAT BD ULTRA FINE¿ PEN NEEDLES HAD NO INSULIN FLOW DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 320122 BATCH NO. 9162522. IT WAS REPORTED THAT THERE WAS NO INSULIN FLOW DURING INJECTION. VERBATIM: CONSUMER STATED, NO INSULIN FLOW WHEN TAKING INJECTION. STATED, SHE DOES NOT PRIME BEFORE INJECTION. STATED, SHE DOES NOT WANT TO WASTE INSULIN PRIMING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 278469 | BD ULTRA FINE¿ PEN NEEDLES | HYPODERMIC SINGLE LUMEN NEEDLE | FMI | BECTON DICKINSON AND CO. | 2865 | 9162522 | 10885403928659 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |