BD ULTRA FINE¿ PEN NEEDLES
Report
- Report Number
- 9616657-2020-00041
- Event Type
- Malfunction
- Date Received
- March 10, 2020
- Date of Event
- February 26, 2020
- Report Date
- February 27, 2020
- Manufacturer
- BECTON, DICKINSON AND CO.
- Product Code
- FMI
- UDI-DI
- 10885403928659
- PMA / PMN Number
- K162516
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. INVESTIGATION SUMMARY: A COMPLAINT HISTORY CHECK WAS PERFORMED AND THIS IS THE 1ST RELATED COMPLAINT FOR NEEDLE CLOG 9171677. NO SAMPLES WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. IF SAMPLES ARE RECEIVED IN THE FUTURE THE COMPLAINT WILL BE REOPENED FOR FURTHER INVESTIGATION. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCE'S WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS AS NO SAMPLES AND/OR PHOTO(S) WERE RECEIVED THE INVESTIGATION CONCLUDED: UNCONFIRMED: BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE AS NO SAMPLES OR PHOTOS WERE RETURNED.
IT WAS REPORTED THAT DURING PRIMING MEDICATION DIDN'T FLOW WITH A BD ULTRA FINE¿ PEN NEEDLES. THIS OCCURRED ON 2 SEPARATE OCCASIONS, HOWEVER, THE DATE/TIME INFORMATION IS UNKNOWN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT CONSUMER FOUND TWO NEEDLES CLOGGED DURING FLOW CHECK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 278500 | BD ULTRA FINE¿ PEN NEEDLES | HYPODERMIC SINGLE LUMEN NEEDLE | FMI | BECTON, DICKINSON AND CO. | 2865 | 9171677 | 10885403928659 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |