FDA Adverse Event
Other
Summary report: N
CUTOM ULTRASONICS SYSTEM 83 PLUS
MDR report key: 980685
·
Received December 30, 2005
Report
- Report Number
- 2523209-2004-00012
- Event Type
- Other
- Date Received
- December 30, 2005
- Date of Event
- April 30, 2004
- Report Date
- September 29, 2004
- Manufacturer
- CUSTOM ULTRASONICS
- Product Code
- FLG
- Report Source
- Manufacturer report
- Reporter Occupation
- INVALID DATA
Narratives
Description of Event or Problem · 1
HOSP STATES O-RINGS WERE SWOLLEN. HOSP STATES THE WERE USING TERGAL 800 AS A DETERGENT BUT RECORDS INDICATE OTHERWISE. (CU TECH) WAS SENT TO THE FACILITY TO CHECK THE SYS AND FOUND THAT ALL SYS (5 TOTAL BAYS) WERE OPERATIONAL TO FACTORY SPECS. CU TECH DID NOTICE IMPROPER CONNECTIONS TO THE SCOPE BY THE OPERATOR OF THE SYS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CUTOM ULTRASONICS SYSTEM 83 PLUS | WASHER DISINFECTOR | FLG | CUSTOM ULTRASONICS | 83+2CI | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Other |