FDA Adverse Event Other Summary report: N

CUSTOM ULTRASONICS SYSTEM 83 PLUS

MDR report key: 980635 · Received December 30, 2005

Report

Report Number
2523209-2004-00005
Event Type
Other
Date Received
December 30, 2005
Date of Event
February 4, 2004
Report Date
September 29, 2004
Manufacturer
CUSTOM ULTRASONICS INC.
Product Code
FLG
Report Source
Manufacturer report
Reporter Occupation
INVALID DATA

Narratives

Description of Event or Problem · 1

HOSP STATES O-RINGS WERE SWOLLEN. HOSP STATES THEY WERE USING TERGAL 800 AS A DETERGENT BUT RECORDS INDICATE OTHERWISE. (CU TECH) WAS SENT TO THE FACILITY TO CHECK THE SYSTEMS AND FOUND THAT ALL SYSTEMS (5 TOTAL BAYS) WHERE OPERATIONAL TO FACTORY SPECS. CU TECH DID NOTICE IMPROPER CONNECTIONS TO THE SCOPE BY THE OPERATOR OF THE SYSTEMS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CUSTOM ULTRASONICS SYSTEM 83 PLUS WASHER DISINFECTOR FLG CUSTOM ULTRASONICS INC. 83+2CI *

Patients

Seq Age Sex Outcome Treatment
1 67 YR Other