Report Number
1213809-2020-00172
Event Type
Malfunction
Date Received
March 9, 2020
Date of Event
February 10, 2020
Report Date
March 18, 2020
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: ONE PHOTO OF A LOOSE 10ML SYRINGE WITH APPROXIMATELY 5ML OF CLEAR FLUID INSIDE WAS RECEIVED AND EVALUATED. IT WAS OBSERVED THERE WAS A LARGE BROWN FOREIGN MATTER PARTICLE FLOATING IN THE PATH. IT APPEARED TO BE A PIECE OF CARDBOARD AND WAS LARGER THAN LEVEL 2 IN SIZE, WHICH WAS REJECTABLE PER PRODUCT SPECIFICATION. POTENTIAL ROOT CAUSE FOR THE FOREIGN MATTER DEFECT IS ASSOCIATED WITH THE MATERIAL HANDLING PROCESS. NO CORRECTIVE ACTIONS ARE NECESSARY BASED ON THE DEFECTIVE RATE IDENTIFIED. BATCH 9242216 IS CONSIDERED IN COMPLIANCE WITH OUR PRODUCT SPECIFICATION REQUIREMENTS. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SYRINGE 10ML LL BNS EXPERIENCED FOREIGN MATTER CONTAMINATION WHICH WAS NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 301029 BATCH NO: 9242216. EVENT DESCRIPTION PER ATTACHED EMAIL STATES: "SYRINGE WITH FOREIGN OBJECT FLOATING IN SALINE."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYRINGE 10ML LL BNS EXPERIENCED FOREIGN MATTER CONTAMINATION WHICH WAS NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 301029 BATCH NO: 9242216. EVENT DESCRIPTION PER ATTACHED EMAIL STATES: "SYRINGE WITH FOREIGN OBJECT FLOATING IN SALINE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
267920 SYRINGE 10ML LL BNS PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 9242216

Patients

Seq Age Sex Outcome Treatment
1 Other