FDA Adverse Event Injury Summary report: N

STEM: AMISTEM H COATED STD STEM SIZE 3

MDR report key: 9805218 · Received March 9, 2020

Report

Report Number
3005180920-2020-00145
Event Type
Injury
Date Received
March 9, 2020
Date of Event
February 12, 2020
Report Date
March 9, 2020
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030804090
PMA / PMN Number
K093944
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 19 FEBRUARY 2020: LOT 123307: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 26-OCT-2012. EXPIRATION DATE: 2017-09-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITH ANOTHER SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

THE PATIENT CAME IN REPORTING OF PAIN DUE TO A POTTED STEM. THE CAUSE OF THE POTTED STEM IS UNKNOWN. THE SURGEON REVISED THE LINER AND STEM 7 YEARS AFTER PRIMARY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
271742 STEM: AMISTEM H COATED STD STEM SIZE 3 HIP UNCEMENTED STEM LZO MEDACTA INTERNATIONAL SA 01.18.133 123307 07630030804090

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention