FDA Adverse Event Malfunction Summary report: N

ULTRASONIC GASTROVIDEOSCOPE

MDR report key: 9804103 · Received March 9, 2020

Doing adverse event research by hand?

The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.

Free to try · every answer cites its records

Report

Report Number
8010047-2020-01695
Event Type
Malfunction
Date Received
March 9, 2020
Report Date
May 14, 2020
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
ODG
PMA / PMN Number
K051541
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION. THE SUBJECT DEVICE HAS NOT BEEN RETURNED TO OMSC BUT WAS RETURNED TO OLYMPUS FRANCE.(OFR). OFR SENT THE DEVICE TO A THIRD PARTY LABORATORY FOR MICROBIOLOGICAL TESTING. AS A RESULT OF THE TESTING, THE SAMPLE COLLECTED FROM THE ALL CHANNELS OF THE DEVICE TESTED POSITIVE FOR COAGULASE-NAGATIVE STAPHYLOCCOCCI (1 CFU/ENDOSCOPE) THE TESTING RESULT CLEARED THE FRENCH GUIDELINE. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION. OMSC REVIEWED THE MANUFACTURING HISTORY (DHR) OF THE SUBJECT DEVICE AND CONFIRMED NO IRREGULARITY. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED.

Description of Event or Problem · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION. THE DEVICE HAD BEEN REPROCESSED WITH A NON-OLYMPUS AUTOMATED ENDOSCOPE REPROCESSOR, WD440, USING PERACETIC ACID. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE IN THIS REPORT HAS NOT BEEN RETURNED TO OMSC FOR EVALUATION. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED.

Description of Event or Problem · 1

OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT AS A RESULT OF MULTIPLE MICROBIOLOGICAL TESTING BY THE USER FACILITY, FOLLOWING MICROBES WERE DETECTED FROM THE SAMPLE COLLECTED FROM THE SUBJECT DEVICE. FIRST TIME; FEBRUARY 5TH, 2020; COAGULASE-NEGATIVE STAPHYLOCCOCCI (>100 CFU/100ML), SECOND TIME; FEBRUARY 7TH, 2020, KOCURIA RHIZOPHILA AND KLEBSIELLA OXYTOCA (THE TOTAL CFU OF THE KOCURIA RHIZOPHILA AND KLEBSIELLA OXYTOCA IS 4 CFU/100ML), THE DEVICE HAD BEEN REPROCESSED USING PERACETIC ACID. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
268460 ULTRASONIC GASTROVIDEOSCOPE ULTRASONIC GASTROVIDEOSCOPE ODG OLYMPUS MEDICAL SYSTEMS CORP. GF-UE160-AL5

Patients

Seq Age Sex Outcome Treatment
1