FDA Adverse Event Malfunction Summary report: N

CADD SOLIS VIP

MDR report key: 9803966 · Received March 8, 2020

Report

Report Number
3012307300-2020-01932
Event Type
Malfunction
Date Received
March 8, 2020
Report Date
April 16, 2020
Manufacturer
SMITHS MEDICAL ASD, INC
Product Code
FRN
UDI-DI
15019517126587
PMA / PMN Number
K111275
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

B5 WAS UPDATED TO REFLECT NO PATIENT INVOLVEMENT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PUMP EXHIBITED ALARM "SETTINGS LOST" WHEN TURNED ON EVERY TIME. NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative · 1

EVALUATION RESULTS: ONE CADD SOLIS VIP WAS RETURNED FOR INVESTIGATION IN USED CONDITION. A REVIEW OF THE EVENT HISTORY LOG EVIDENCED THE FOLLOWING: "(B)(6) 2005 12:00:07 AM MEDIUM ALARM DISPLAYED: PUMP SETTINGS AND PATIENT DATA LOST. (B)(6) 2005 12:00:12 AM MEDIUM ALARM ACKNOWLEDGED: PUMP SETTINGS AND PATIENT DATA LOST. (B)(6) 2005 12:00:15 AM INFO ALARM: USB CABLE ATTACHED." THE CUSTOMER REPORTED PRODUCT PROBLEM WAS REPLICATED. THE ERROR "PUMP SETTINGS AND PATIENT DATA LOST" WAS DUPLICATED AT POWER UP. THE PRODUCT PROBLEM WAS ATTRIBUTED TO THE INTERNAL BACKUP BATTERY BEEN DRAINED OUT. THE CUSTOMER MAY HAVE LEFT THE PUMP SITTING IN THE STORAGE FOR A LONG PERIOD OF TIME WITHOUT POWERING UP THE DEVICE. THE PUMP WAS POWERED FOR AN EXTENDED PERIOD OF TIME SO THE INTERNAL BACKUP BATTERY COULD CHARGE. NO PARTS WERE REPLACED. THE PROBLEM SOURCE OF THE REPORTED PRODUCT PROBLEM WAS NOT DEFINITIVELY KNOWN. A ROOT CAUSE WAS THERE NOT ESTABLISHED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PUMP EXHIBITED ALARM "SETTINGS LOST" WHEN TURNED ON EVERY TIME. NO PATIENT INJURY OR COMPLICATIONS WERE REPORTED IN RELATION TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
267330 CADD SOLIS VIP PUMP, INFUSION FRN SMITHS MEDICAL ASD, INC 2120 15019517126587

Patients

Seq Age Sex Outcome Treatment
1