AIA-360
Report
- Report Number
- 8031673-2020-00045
- Event Type
- Malfunction
- Date Received
- March 7, 2020
- Date of Event
- February 10, 2020
- Report Date
- March 18, 2020
- Manufacturer
- TOSOH CORPORATION
- Product Code
- KHO
- PMA / PMN Number
- K971103
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT THE MANUFACTURER'S PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT. FSE ARRIVED ONSITE TO ADDRESS THE REPORTED EVENT. FSE CONFIRMED THE ISSUE BY REVIEWING THE CUSTOMER'S PRINTOUTS AND REPRODUCED THE ISSUE BY RUNNING A PATIENT SAMPLE. NEXT, FSE REPLACED THE WASH PROBE AND VERIFIED ALIGNMENTS WERE WITHIN SPECIFICATIONS. FSE ALSO DECONTAMINATED THE SYSTEM DUE TO SUSPECTED CONTAMINATION AND UPGRADED THE SOFTWARE PER THE T187 BULLETIN. NO FURTHER ISSUES WERE NOTED. NO FURTHER ACTION AS REQUIRED BY FIELD SERVICE. THE MOST PROBABLE CAUSE OF THE REPORTED EVENT WAS DUE TO FAILURE OF THE WASH PROBE AND SYSTEM CONTAMINATION.
INITIALLY, FSE WORKED WITH THE CUSTOMER OVER THE PHONE TO PERFORM DECONTAMINATION OF THE ANALYZER. THE CUSTOMER RAN E2 WITH DI AND RECEIVED THE RESULT 74.2PG/M. NEXT, THE CUSTOMER SENT IN THE FROZEN SAMPLES TO BE TESTED IN THE QUALITY ASSURANCE (QA) LAB. SEVERAL DAYS LATER FSE ARRANGED TO GO ONSITE AS THE ISSUE WAS STILL ONGOING. A 13-MONTH COMPLAINT SERVICE HISTORY REVIEW FOR SIMILAR COMPLAINTS WAS PERFORMED FOR SERIAL NUMBER (B)(4) FROM 10JAN2019 THROUGH AWARE DATE 10FEB2020. THERE WAS TWO SIMILAR COMPLAINTS, INCLUDING THIS ONE, IDENTIFIED DURING THE SEARCHED PERIOD. A 13- MONTH LOT-SERIAL NUMBER SPECIFIC HISTORY REVIEW WAS PERFORMED FOR E2 LOT J912944 FROM 10JAN2019 THROUGH AWARE DATE 10FEB2020. THERE WERE NO SIMILAR COMPLAINTS IDENTIFIED DURING THE SEARCHED PERIOD. THE ST AIA-PACK E2 ANALYTE APPLICATION MANUAL STATES THE FOLLOWING: LIMITATIONS OF THE PROCEDURE: FOR DIAGNOSTIC PURPOSES, THE RESULTS OBTAINED FROM THIS ASSAY SHOULD BE USED IN CONJUNCTION WITH OTHER DATA (E.G. SYMPTOMS, RESULTS OF OTHER TESTS, CLINICAL IMPRESSIONS, THERAPY, ETC.). EXPECTED VALUES: EACH LABORATORY SHOULD DETERMINE A REFERENCE INTERVAL CORRESPONDING TO THE CHARACTERISTICS OF THE POPULATION BEING TESTED. AS WITH ALL DIAGNOSTIC PROCEDURES, CLINICAL RESULTS MUST BE INTERPRETED WITH REGARD TO CONCOMITANT MEDICATIONS ADMINISTERED TO THE PATIENT. THE MOST PROBABLE CAUSE OF THE REPORTED EVENT IS STILL BEING INVESTIGATED.
THE CUSTOMER REPORTED RECEIVING DISCREPANT RESULTS ON TWO PATIENT SAMPLES FOR ESTRADIOL (E2) ON THEIR AIA-360 ANALYZER. THE RESULT FOR ONE SAMPLE WAS 64 AND 68.3 ON THE OTHER USING E2 LOT # J912944 AND CALIBRATOR LOT #J632989. THE CUSTOMER ALSO SENT THE SAME SAMPLE OUT TO A REFERENCE LAB AND RECEIVED RESULTS OF <0.05 FOR BOTH. THE CUSTOMER STATED THAT THE SAMPLES WERE NOT FROZEN PRIOR TO SENDING THEM TO THE REFERENCE LAB. THE CUSTOMER ALSO STATED THAT NEITHER PATIENT WAS ON ANY OTHER SUPPLEMENT OTHER THAN THE ORAL E2 INHIBITOR SO BIOTIN INTERFERENCE WAS RULED OUT. THESE WERE THE ONLY TWO PATIENTS REPORTED. THERE WAS NO SIMILAR ISSUE WITH ANY OTHER ANALYTE. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ADDRESS THE REPORTED EVENT. THERE WAS NO INDICATION OF ANY PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE REPORTED EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 266980 | AIA-360 | AIA-360 | KHO | TOSOH CORPORATION | AIA-360 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |