FDA Adverse Event Other Summary report: N

CUSTOM ULTRASONICS SYSTEMS 83 PLUS

MDR report key: 980242 · Received December 30, 2005

Report

Report Number
2523209-2004-00003
Event Type
Other
Date Received
December 30, 2005
Date of Event
December 9, 2003
Report Date
September 29, 2004
Manufacturer
CUSTOM ULTRASONICS, INC
Product Code
FLG
Report Source
Manufacturer report
Reporter Occupation
INVALID DATA

Narratives

Description of Event or Problem · 1

HOSP STATES O-RINGS WERE SWOLLEN. HOSP STATES THE WERE USING TERGAL 800 AS A DETERGENT BUT RECORDS INDICATE OTHERWISE. (CU TECH) WAS SENT TO THE FACILITY TO CHECK THE SYS AND FOUND THAT ALL SYS (5 TOTAL BAYS) WERE OPERATIONAL TO FACTORY SPECS. CU TECH DID NOTICE IMPROPER CONNECTIONS TO THE SCOPE BY THE OPERATOR OF THE SYS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CUSTOM ULTRASONICS SYSTEMS 83 PLUS WASHER DISINFECTOR FLG CUSTOM ULTRASONICS, INC 83-21CI *

Patients

Seq Age Sex Outcome Treatment
1 45 YR Other