COATED BIO-EYE HA IMPLANT WITH COMFORMER
Report
- Report Number
- 2027377-2007-00010
- Event Type
- Injury
- Date Received
- January 15, 2008
- Date of Event
- December 18, 2007
- Report Date
- December 18, 2007
- Manufacturer
- INTEGRATED ORBITAL IMPLANTS, INC.
- Product Code
- HPZ
- PMA / PMN Number
- K003338
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PHYSICIAN
Narratives
EXPOSURE IS AN ANTICIPATED COMPLICATION OF ORBITAL IMPLANT SURGERY. IT IS GENERALLY TREATED CONSERVATIVELY AND IS SURGICALLY CLOSED ONLY IN THE CASE OF LARGE EXPOSURES OR IF THE WOUND DOES NOT SPONTANEOUSLY CLOSE WITHIN A REASONABLE TIME PERIOD. THE ANTERIOR SHELL OF THE COATED BIO-EYE IMPLANT IS DESIGNED TO RESORB IN 12-18 MONTHS. LESS THAN 15 MONTHS HAD PASSED SINCE IMPLANTATION. DISSOLUTION RATES OF BIORESORBABLE POLYMERS IS DEPENDENT UPON A NUMBER OF CHARACTERISTICS INCLUDING TEMPERATURES AND HYDRATION, AND IS ANTICIPATED TO VARY BASED UPON THE INDIVIDUAL PATIENT'S PHYSIOLOGY. IN THE EVENT THAT THE EXPOSURE WAS OF PROLONGED DURATION THE POLYMER WOULD NOT BE HYDRATED AND THEREFORE NOT SUBJECT TO THE DESIGNED RATE OF DEGRADATION. DR. RETURNED SAMPLES OF THE PIECES OF COATING SHE REMOVED. THESE WERE EVALUATED AS FOLLOWS: THE EVALUATION CONSISTED OF VISUAL AND PHOTOGRAPHIC EVALUATION AS WELL AS THICKNESS MEASUREMENTS. TO DATE WE HAVE NO VIABLE HYPOTHESIS AS TO THE GENESIS OF THE DISTINCTIVE SHAPES OF THE FRAGMENTS. THE FRAGMENTS PROVIDED MAY HAVE BEEN CUT INTO THEIR CHARACTERISTICS SHAPES BY THE ACTION OF THE PHYSICIAN IN REMOVING THE EXPOSED POLYMER.
DR. REPORTED THAT THE PATIENT HAD BLEEDING AND IRRITATION TO HER EYE PERHAPS CAUSED BY THE COATED BIO-EYE IMPLANT. DR. OPENED UP THE EYE AND FOUND SHARP EDGES OF IMPLANT POKING THROUGH CONJUNCTIVA. SHE CLEANED THE ANTERIOR PART AND "NIPPED AWAY" ABOUT 20-30% OF IMPLANT. SHE TOOK A CULTURE AND CLOSED THE CONJUNCTIVA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COATED BIO-EYE HA IMPLANT WITH COMFORMER | IMPLANT, EYE SPHERE | HPZ | INTEGRATED ORBITAL IMPLANTS, INC. | I0018C | 36307 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Required Intervention |