FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND HIGH PROFILE

MDR report key: 9795423 · Received March 5, 2020

Report

Report Number
1645337-2020-03709
Event Type
Injury
Date Received
March 5, 2020
Date of Event
February 8, 2020
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001645
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION OF THE RETURNED DEVICE WAS COMPLETED BY THE FAILURE ANALYSIS LAB ON (B)(6) 2020. DEVICE EVALUATION SUMMARY: ACCORDING TO THE INFORMATION PROVIDED, THE PATIENT EXPERIENCED A DEFLATION OF THE IMPLANT. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED FOR THE FINISHED DEVICE NUMBER 6844055, AND NO NON-CONFORMANCE'S RELATED TO THE REPORTED COMPLAINT WERE IDENTIFIED. THE DEVICE WAS VISUALLY INSPECTED, AND A TEAR MEASURING APPROXIMATELY 1.5 CM WAS NOTED ON THE POSTERIOR ASPECT. MICROSCOPIC EXAMINATION OF THE EDGES OF THE RUPTURE REVEALED PARALLEL STRIATIONS CONSISTENT WITH A RUPTURE WITH A SHARP OBJECT. THE STRIATIONS ARE CONSISTENT WITH MARKINGS MADE BY A SHARP INSTRUMENT PERFORATING SILICONE MATERIAL. AS STATED IN THE PRODUCT INSERT DATA SHEET, DO NOT ALLOW CAUTERY DEVICES OR SHARP INSTRUMENTS, SUCH AS SCALPELS, SUTURE NEEDLES, HYPODERMIC NEEDLES, HEMOSTATS, ADSON FORCEPS OR SCISSORS TO CONTACT THE DEVICE DURING THE IMPLANTATION OR OTHER SURGICAL PROCEDURES AS THIS CAN RESULT IN RUPTURE. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. A MANUFACTURING RECORD EVALUATION IS IN PROGRESS. ONCE COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: DEFLATION. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) YEAR-OLD CAUCASIAN FEMALE WHO UNDERWENT PRIMARY BREAST AUGMENTATION WITH A MENTOR SMOOTH ROUND HIGH PROFILE 330CC SALINE PROSTHESIS EXPERIENCED SPONTANEOUS RIGHT SIDED DEFLATION POST PROCEDURE. THE PATIENT VISITED THE PHYSICIAN¿S OFFICE AND RECEIVED A MEDICAL DIAGNOSIS FOR THE DEFLATION. AS A RESULT, BILATERAL PROSTHESIS REPLACEMENT WITH NEW MENTOR SMOOTH ROUND HIGH PROFILE 330CC SALINE PROSTHESES WAS PERFORMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
258575 MENTOR SMOOTH ROUND HIGH PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 3503330 6844055 00081317001645

Patients

Seq Age Sex Outcome Treatment
1 36 YR Required Intervention