FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM

MDR report key: 9788327 · Received March 4, 2020

Report

Report Number
3006948883-2020-00075
Event Type
Malfunction
Date Received
March 4, 2020
Date of Event
February 14, 2020
Report Date
April 6, 2020
Manufacturer
BD (SUZHOU)
Product Code
FOZ
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 9170827. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. ADDITIONALLY, A SAMPLE WAS RECEIVED FOR THE PURPOSE OF OUR INVESTIGATION. OUR INVESTIGATORS NOTED A SMALL CRACK WAS FOUND DURING LEAKAGE TESTING THAT WAS LOCATED ON THE ADAPTER. THE RETURNED PRODUCT'S SWAGE DIMENSIONS WERE MEASURED BY OUR TEAM AND FOUND TO BE WITHIN PRODUCT SPECIFICATIONS. HOWEVER ACCORDING TO PREVIOUS INVESTIGATIONS, THROUGH A VARIANCE IN THE AUTOLINE'S SWAGE PRESSURE IT IS POSSIBLE FOR THE MACHINE TO APPLY EXCESS FORCE TO THE DEVICE DURING ASSEMBLY, ALLOWING FOR THE RESULTING CRACK TO FORM IN THE DEVICE. TO ADDRESS THIS ISSUE THE MOLDING CAVITY HAS BEEN OPTIMIZED TO INCREASE RESILIENCE OF THE ADAPTER DURING THE SWAGING PROCESS . H3 OTHER TEXT : SEE SECTION H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM LEAKED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LEAKAGE WAS FOUND IN THE CATHETER SEAT PLACE DURING THE USE OF THE PRODUCT.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM LEAKED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LEAKAGE WAS FOUND IN THE CATHETER SEAT PLACE DURING THE USE OF THE PRODUCT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
253304 BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM CATHETER FOZ BD (SUZHOU) 9170827

Patients

Seq Age Sex Outcome Treatment
1 Other