Report

Report Number
1911916-2020-00211
Event Type
Malfunction
Date Received
March 4, 2020
Date of Event
February 13, 2020
Report Date
March 6, 2020
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
UDI-DI
30382903028321
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: 1 SAMPLE WAS RECEIVED. NO PHOTO WAS PROVIDED. ONE SAMPLE WAS RECEIVED. IT CAME IN AN OPEN PACKAGING BLISTER. VISUAL INSPECTION WAS PERFORMED. IT HAS THE PLUNGER ROD THUMB PRESS DAMAGED. NO DEFECTS OR DAMAGES WERE OBSERVED IN ANY OTHER AREA. THIS WOULD HAVE HAPPENED DURING SYRINGE ASSEMBLY PROCESS. A DEVICE HISTORY REVIEW COULD NOT BE COMPLETED AS NO BATCH NUMBER WAS PROVIDED.

Description of Event or Problem · 0

IT WAS REPORTED THAT 30 ML BD LUER-LOK¿ TIP SYRINGE WAS BROKEN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO.: 302832 BATCH NO.: UNKNOWN (PROVIDED 924053). IT WAS REPORTED THE TOP OF THE SYRINGE WAS BROKEN OFF CREATING A VERY SHARP EDGE. VERBATIM: TOP OF SYRINGE WAS BROKEN OFF CREATING A VERY SHARP EDGE. THE NURSE ALMOST CUT HER HAND ON IT.

Additional Manufacturer Narrative · 1

MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN.

Description of Event or Problem · 1

IT WAS REPORTED THAT 30 ML BD LUER-LOK¿ TIP SYRINGE WAS BROKEN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO.: 302832 BATCH NO.: UNKNOWN (PROVIDED 924053). IT WAS REPORTED THE TOP OF THE SYRINGE WAS BROKEN OFF CREATING A VERY SHARP EDGE. VERBATIM: TOP OF SYRINGE WAS BROKEN OFF CREATING A VERY SHARP EDGE. THE NURSE ALMOST CUT HER HAND ON IT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
253436 30 ML BD LUER-LOK TIP SYRINGE PISTON SYRINGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 302832 UNKNOWN 30382903028321

Patients

Seq Age Sex Outcome Treatment
1 Other