FDA Adverse Event Injury Summary report: N

SMARTSET GHV GENTAMICIN 40G

MDR report key: 9781791 · Received March 3, 2020

Report

Report Number
1818910-2020-06726
Event Type
Injury
Date Received
March 3, 2020
Date of Event
September 5, 2019
Report Date
September 5, 2019
Manufacturer
DEPUY CMW - 9610921
Product Code
LOD
UDI-DI
10603295174288
PMA / PMN Number
K033563
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY : NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT : A PREVIOUS DHR REVIEW (B)(4) DID NOT REVEAL ANY RELATED MANUFACTURING DEVIATIONS OR ANOMALIES ON THE REPORTED LOT NUMBER (8053415).

Additional Manufacturer Narrative · 1

(B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

THE PATIENT WAS REVISED TO ADDRESS PAIN AND LOOSENING OF THE TIBIAL COMPONENT AT THE CEMENT TO IMPLANT INTERFACE. CEMENT MANUFACTURER IS UNKNOWN. ON (B)(6) 2015: PATIENT UNDERWENT A PRIMARY KNEE REPLACEMENT. DEPUY IMPLANTS WERE IMPLANTED WITH DEPUY CEMENT X2. NO INTRA-OPERATIVE COMPLICATIONS WERE NOTED. ON (B)(6) 2019: PATIENT PRESENTED WITH PAIN AND UNDERWENT A RIGHT KNEE REVISION. THE TIBIAL TRAY WAS NOTED TO BE IN VARUS AND LOOSE AT THE IMPLANT TO CEMENT INTERFACE. THE FEMORAL COMPONENT AND PATELLA WERE NOTED TO BE WELL FIXED AND RETAINED. ATTUNE REVISION IMPLANTS WERE USED WITH DEPUY CEMENT. DOI: UNKNOWN. DOR: (B)(6) 2019, RIGHT KNEE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
244178 SMARTSET GHV GENTAMICIN 40G BONE CEMENT : BONE CEMENT LOD DEPUY CMW - 9610921 5450-35-500 8053415 10603295174288

Patients

Seq Age Sex Outcome Treatment
1 45 YR Required Intervention ATTUNE MEDIAL DOME PAT 38MM| ATTUNE PS FEM RT SZ 5 CEM| ATTUNE PS RP INSRT SZ 5 7MM| ATTUNE RP TIB BASE SZ 5 CEM| SMARTSET GHV GENTAMICIN 40G