FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SODIUM FLUORIDE POTASSIUM OXALATE (FX) BLOOD COLLECTION TUBES

MDR report key: 9781445 · Received March 3, 2020

Report

Report Number
1917413-2020-00219
Event Type
Malfunction
Date Received
March 3, 2020
Date of Event
February 12, 2020
Report Date
March 31, 2020
Manufacturer
BECTON, DICKINSON & CO. (BROKEN BOW)
Product Code
JKA
UDI-DI
50382903679220
PMA / PMN Number
SEE H.10.
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: BD HAD NOT RECEIVED SAMPLES, BUT PHOTOS WERE PROVIDED BY THE CUSTOMER FACILITY FOR INVESTIGATION. THE PHOTOS WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE FOR LABEL LIFT WITH THE INCIDENT LOT WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCE'S DURING MANUFACTURING OF THE PRODUCT. BD HAS INITIATED FURTHER INVESTIGATION RELATING TO THIS ISSUE THROUGH CAPA#1064141 AND POTENTIAL CAUSES HAVE BEEN IDENTIFIED. AS A RESULT, CORRECTIVE ACTIONS HAVE BEEN ESTABLISHED AND ARE IN THE PROCESS OF BEING IMPLEMENTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT 50 BD VACUTAINER® SODIUM FLUORIDE POTASSIUM OXALATE (FX) BLOOD COLLECTION TUBES FROM LOT# 9260611, AND 140 TUBES FROM LOT# 9220487, HAD ISSUES WITH LABEL PEEL OFF BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "OPEN LABEL LOT.9260611 = 50 SP / LOT.9220487 = 140 SP".

Additional Manufacturer Narrative · 1

THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 9260611. MEDICAL DEVICE EXPIRATION DATE: 2021-01-31. DEVICE MANUFACTURE DATE: 2019-09-17. MEDICAL DEVICE LOT #: 9220487. MEDICAL DEVICE EXPIRATION DATE: 2020-12-31. DEVICE MANUFACTURE DATE: 2019-08-08. THERE WERE MULTIPLE PMA / 510(K)#S REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH 510(K) NUMBER IS AS FOLLOWS: PMA / 510(K)#: K945952, PMA / 510(K)#: K901449. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT 50 BD VACUTAINER® SODIUM FLUORIDE POTASSIUM OXALATE (FX) BLOOD COLLECTION TUBES FROM LOT# 9260611, AND 140 TUBES FROM LOT# 9220487, HAD ISSUES WITH LABEL PEEL OFF BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "OPEN LABEL LOT.9260611 = 50 SP / LOT.9220487 = 140 SP."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
246536 BD VACUTAINER® SODIUM FLUORIDE POTASSIUM OXALATE (FX) BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO. (BROKEN BOW) 367922 SEE SECTION H.10. 50382903679220

Patients

Seq Age Sex Outcome Treatment
1 Other