BD VACUTAINER® SODIUM FLUORIDE POTASSIUM OXALATE (FX) BLOOD COLLECTION TUBES
Report
- Report Number
- 1917413-2020-00219
- Event Type
- Malfunction
- Date Received
- March 3, 2020
- Date of Event
- February 12, 2020
- Report Date
- March 31, 2020
- Manufacturer
- BECTON, DICKINSON & CO. (BROKEN BOW)
- Product Code
- JKA
- UDI-DI
- 50382903679220
- PMA / PMN Number
- SEE H.10.
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION: BD HAD NOT RECEIVED SAMPLES, BUT PHOTOS WERE PROVIDED BY THE CUSTOMER FACILITY FOR INVESTIGATION. THE PHOTOS WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE FOR LABEL LIFT WITH THE INCIDENT LOT WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCE'S DURING MANUFACTURING OF THE PRODUCT. BD HAS INITIATED FURTHER INVESTIGATION RELATING TO THIS ISSUE THROUGH CAPA#1064141 AND POTENTIAL CAUSES HAVE BEEN IDENTIFIED. AS A RESULT, CORRECTIVE ACTIONS HAVE BEEN ESTABLISHED AND ARE IN THE PROCESS OF BEING IMPLEMENTED.
IT WAS REPORTED THAT 50 BD VACUTAINER® SODIUM FLUORIDE POTASSIUM OXALATE (FX) BLOOD COLLECTION TUBES FROM LOT# 9260611, AND 140 TUBES FROM LOT# 9220487, HAD ISSUES WITH LABEL PEEL OFF BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "OPEN LABEL LOT.9260611 = 50 SP / LOT.9220487 = 140 SP".
THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 9260611. MEDICAL DEVICE EXPIRATION DATE: 2021-01-31. DEVICE MANUFACTURE DATE: 2019-09-17. MEDICAL DEVICE LOT #: 9220487. MEDICAL DEVICE EXPIRATION DATE: 2020-12-31. DEVICE MANUFACTURE DATE: 2019-08-08. THERE WERE MULTIPLE PMA / 510(K)#S REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH 510(K) NUMBER IS AS FOLLOWS: PMA / 510(K)#: K945952, PMA / 510(K)#: K901449. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT 50 BD VACUTAINER® SODIUM FLUORIDE POTASSIUM OXALATE (FX) BLOOD COLLECTION TUBES FROM LOT# 9260611, AND 140 TUBES FROM LOT# 9220487, HAD ISSUES WITH LABEL PEEL OFF BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "OPEN LABEL LOT.9260611 = 50 SP / LOT.9220487 = 140 SP."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 246536 | BD VACUTAINER® SODIUM FLUORIDE POTASSIUM OXALATE (FX) BLOOD COLLECTION TUBES | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON & CO. (BROKEN BOW) | 367922 | SEE SECTION H.10. | 50382903679220 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |