FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® BUFFERED SODIUM CITRATE (9NC) BLOOD COLLECTION TUBES

MDR report key: 9781221 · Received March 3, 2020

Report

Report Number
1917413-2020-00218
Event Type
Malfunction
Date Received
March 3, 2020
Date of Event
February 12, 2020
Report Date
April 21, 2020
Manufacturer
BECTON, DICKINSON & CO. (BROKEN BOW)
Product Code
JKA
UDI-DI
50382903630832
PMA / PMN Number
K013971
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY BD HAD NOT RECEIVED SAMPLES OR PHOTOS FROM THE CUSTOMER FACILITY FOR EVALUATION. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE TUBES ARE OVER FILLING WITH A BD VACUTAINER® BUFFERED SODIUM CITRATE (9NC) BLOOD COLLECTION TUBES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT TUBES ARE OVERFILLING PAST THE LINE.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. "MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 9220391. MEDICAL DEVICE EXPIRATION DATE: 2020-05-31. DEVICE MANUFACTURE DATE: 2019-08-08. MEDICAL DEVICE LOT #: 9260286. MEDICAL DEVICE EXPIRATION DATE: 2020-06-30. DEVICE MANUFACTURE DATE: 2019-09-17." A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING USE TUBES ARE OVER FILLING WITH A BD VACUTAINER® BUFFERED SODIUM CITRATE (9NC) BLOOD COLLECTION TUBES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT TUBES ARE OVERFILLING PAST THE LINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
246840 BD VACUTAINER® BUFFERED SODIUM CITRATE (9NC) BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO. (BROKEN BOW) 363083 SEE H.10 50382903630832

Patients

Seq Age Sex Outcome Treatment
1 Other