MCKESSON BRANDS
Report
- Report Number
- 1451040-2020-00002
- Event Type
- Injury
- Date Received
- March 3, 2020
- Date of Event
- February 19, 2020
- Report Date
- March 2, 2020
- Manufacturer
- SHANGHAI JIANZHONG MEDICAL PACKAGING CO, LTD
- Product Code
- FRG
- Adverse Event
- Yes
- Report Source
- Distributor report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
CUSTOMER REPORTED THAT THEIR SELF SEAL STERILIZATION POUCHES CAUGHT FIRE IN THEIR RITTER M7 SPEEDCLAVE AUTOCLAVE (SERIAL # (B)(4)). UPON OPENING THE MACHINE SHE NOTICED THAT THE BAGS WERE BURNING. SHE THEN PUT THE AUTOCLAVE ON A CART AND TOOK IT OUTSIDE. THE AUTOCLAVE WAS REPORTED TO BE 10 YEARS OLD. MCKESSON CONTACTED THE MANUFACTURER OF THE AUTOCLAVE TO BRING THIS TO THEIR ATTENTION AS WELL. THEY INDICATED THAT THE UNIT WAS BUILT AND SHIPPED IN 2003 (17 YEARS AGO). THEY INDICATED THAT THERE IS POTENTIAL FOR THE OPERATOR TO TURN THE UNIT ON WITHOUT WATER IN THE CHAMBER, WHICH WOULD RESULT IN THE CHAMBER GETTING VERY HOT. THERE IS ALSO THE POTENTIAL FOR A VALVE OR GASKET TO NOT SEAL COMPLETELY AND ALLOW THE WATER THAT WAS INITIALLY IN THE CHAMBER TO ESCAPE AND ALLOW THE CHAMBER TO GET VERY HOT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 241814 | MCKESSON BRANDS | POUCH, SELF SEAL ECON 3.5X9 | FRG | SHANGHAI JIANZHONG MEDICAL PACKAGING CO, LTD | CJZI11-02 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |