FDA Adverse Event Injury Summary report: N

MCKESSON BRANDS

MDR report key: 9779313 · Received March 3, 2020

Report

Report Number
1451040-2020-00002
Event Type
Injury
Date Received
March 3, 2020
Date of Event
February 19, 2020
Report Date
March 2, 2020
Manufacturer
SHANGHAI JIANZHONG MEDICAL PACKAGING CO, LTD
Product Code
FRG
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THEIR SELF SEAL STERILIZATION POUCHES CAUGHT FIRE IN THEIR RITTER M7 SPEEDCLAVE AUTOCLAVE (SERIAL # (B)(4)). UPON OPENING THE MACHINE SHE NOTICED THAT THE BAGS WERE BURNING. SHE THEN PUT THE AUTOCLAVE ON A CART AND TOOK IT OUTSIDE. THE AUTOCLAVE WAS REPORTED TO BE 10 YEARS OLD. MCKESSON CONTACTED THE MANUFACTURER OF THE AUTOCLAVE TO BRING THIS TO THEIR ATTENTION AS WELL. THEY INDICATED THAT THE UNIT WAS BUILT AND SHIPPED IN 2003 (17 YEARS AGO). THEY INDICATED THAT THERE IS POTENTIAL FOR THE OPERATOR TO TURN THE UNIT ON WITHOUT WATER IN THE CHAMBER, WHICH WOULD RESULT IN THE CHAMBER GETTING VERY HOT. THERE IS ALSO THE POTENTIAL FOR A VALVE OR GASKET TO NOT SEAL COMPLETELY AND ALLOW THE WATER THAT WAS INITIALLY IN THE CHAMBER TO ESCAPE AND ALLOW THE CHAMBER TO GET VERY HOT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
241814 MCKESSON BRANDS POUCH, SELF SEAL ECON 3.5X9 FRG SHANGHAI JIANZHONG MEDICAL PACKAGING CO, LTD CJZI11-02

Patients

Seq Age Sex Outcome Treatment