FDA Adverse Event Injury Summary report: N

GDC 18-SOFT SYNERG DETECTION CIRCUIT

MDR report key: 977889 · Received January 10, 2008

Report

Report Number
2939204-2008-00006
Event Type
Injury
Date Received
January 10, 2008
Date of Event
December 8, 2007
Report Date
December 11, 2007
Manufacturer
BOSTON SCIENTIFIC CORP.
Product Code
HCG
PMA / PMN Number
K991134
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL K 001083. THE ROOT CAUSE FOR THIS COMPLAINT CANNOT BE DETERMINED DEFINITIVELY. THE RISK OF STRETCHING WHILE MANEUVERING A DETACHABLE COIL IS NOTED IN THE DFU (DIRECTIONS FOR USE) FOR THE DEVICE IN QUESTION: "DUE TO THE DELICATE NATURE OF THE GDC COILS, ...., A COIL MAY OCCASIONALLY STRETCH WHILE BEING MANEUVERED. STRETCHING IS A PRECURSOR TO POTENTIAL MALFUNCTIONS, SUCH AS COIL BREAKAGE." A POSSIBLE CONTRIBUTORY FACTOR TO COIL BREAKAGE MAY BE DUE TO REPOSITIONING OF THE COIL. THE FOLLOWING IS STATED AS A PRECAUTION IN THE DFU: "IF COIL REPOSITIONING IS NECESSARY, TAKE SPECIAL CARE TO RETRACT COIL UNDER FLUOROSCOPY IN A ONE-TO-ONE MOTION WITH THE DELIVERY WIRE. IF THE COIL DOES NOT MOVE WITH A ONE-TO-ONE MOTION, OR REPOSITIONING IS DIFFICULT, THE COIL HAS BEEN STRETCHED AND COULD POSSIBLY BREAK."

Description of Event or Problem · 1

IT WAS REPORTED TO THE MANUFACTURER THAT A CUSTOMER ENCOUNTERED DIFFICULTIES DURING USE OF A DETACHABLE COIL. ACCORDING TO THE CUSTOMER: THEY ADVANCED THROUGH A PROGRADE NEURO CATHETER AND THE COIL [DEVICE IN QUESTION] WAS REPOSITIONED SEVERAL TIMES; THE COIL BECAME STRETCHED AND THEN BROKE. THE PHYSICIAN WAS ABLE TO RETRIEVE THE COIL PIECES USING A GOOSE NECK SNARE, AND ANOTHER COIL WAS SUCCESSFULLY PLACED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED AND THE PT'S POST PROCEDURE CONDITION STATUS WAS REPORTED AS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GDC 18-SOFT SYNERG DETECTION CIRCUIT HCG: NEUROVASCULAR EMBOLIZATION COIL HCG BOSTON SCIENTIFIC CORP. 351512-4 9569548

Patients

Seq Age Sex Outcome Treatment
1 YR Required Intervention