FDA Adverse Event Malfunction Summary report: N

BD PEN NDL 32G 4MM HP

MDR report key: 9770662 · Received February 28, 2020

Report

Report Number
9616656-2020-00169
Event Type
Malfunction
Date Received
February 28, 2020
Date of Event
February 5, 2020
Report Date
March 9, 2020
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FMI
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY THIRTY FIVE SEALED AND INTACT 32G X 4MM PEN NEEDLE SAMPLES WERE RETURNED FROM LOT. NO. 9141583, CAT. NO. 320561. A CLOG TEST WAS CARRIED OUT ON ALL THIRTY FIVE SAMPLES AS PER TP700483 AND NO ISSUES WERE OBSERVED. INVESTIGATION CONCLUSION A CLOG TEST WAS CARRIED OUT ON ALL THIRTY FIVE SAMPLES AS PER TP700483 AND NO ISSUES WERE OBSERVED. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. ROOT CAUSE DESCRIPTION NO ISSUES WERE OBSERVED WITH THE RETURNED SAMPLES THEREFORE NO ROOT CAUSE CAN BE IDENTIFIED. RATIONALE BASED ON AN EVALUATION OF SEVERITY AND OCCURRENCE IT WAS DETERMINED THAT NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME.

Description of Event or Problem · 0

IT WAS REPORTED THAT INSULIN LEAKS FROM SIDE OF PEN DURING USE WITH A BD PEN NDL 32G 4MM HP. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM GERMAN TO ENGLISH: INSULIN LEAKS FROM THE SIDE OF THE PEN.

Additional Manufacturer Narrative · 1

INITIAL REPORTER PHONE #: UNKNOWN. A SAMPLE IS AVAILABLE FOR EVALUATION. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. .

Description of Event or Problem · 1

IT WAS REPORTED THAT INSULIN LEAKS FROM SIDE OF PEN DURING USE WITH A BD PEN NDL 32G 4MM HP. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: INSULIN LEAKS FROM THE SIDE OF THE PEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
235727 BD PEN NDL 32G 4MM HP HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON AND CO. 9141583

Patients

Seq Age Sex Outcome Treatment
1 Other