FDA Adverse Event Malfunction Summary report: N

SYRINGE 50ML LL CLEAR 14GA 1-1/4IN

MDR report key: 9767818 · Received February 28, 2020

Report

Report Number
3002682307-2020-00070
Event Type
Malfunction
Date Received
February 28, 2020
Date of Event
February 1, 2020
Report Date
April 10, 2020
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY THREE SAMPLES WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THE PRODUCT WAS VISUALLY INSPECTED, THE SCALE ON TWO OF THE SYRINGES WAS OBSERVED TO BE MISSING, THE THIRD SAMPLE HAD LIGHT LINES BUT THE SCALE WAS NOT LEGIBLE. A DEVICE HISTORY RECORD REVIEW DID NOT REVEAL ANY DOCUMENTED QUALITY ISSUES DURING THE PRODUCTION OF LOT NUMBER 1904243 THAT COULD HAVE CONTRIBUTED TO THIS INCIDENT. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURE. WHILE WE COULD NOT IDENTIFY A DIRECT ISSUE, IT WAS DETERMINED THIS INCIDENT LIKELY OCCURRED DUE A FAILURE IN THE TRANSFER OF INK FROM THE DRUM TO THE BARREL DURING THE MARKING PROCESS. COMPLAINTS RECEIVED FOR THIS DEFECT AND DEVICE WILL BE MONITORED BY OUR QUALITY TEAM FOR SIGNS OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF SYRINGES 50ML LL CLEAR 14GA 1-1/4IN EXPERIENCED LIGHT/MISSING SCALE MARKINGS. PRODUCT DEFECTS WERE NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THERE IS MISSING SCALE ON SOME OF THE SYRINGES OR THE SCALE IS ONLY WEAKLY PRINTED.

Additional Manufacturer Narrative · 1

(B)(6). A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF SYRINGES 50ML LL CLEAR 14GA 1-1/4IN EXPERIENCED LIGHT/MISSING SCALE MARKINGS. PRODUCT DEFECTS WERE NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THERE IS MISSING SCALE ON SOME OF THE SYRINGES OR THE SCALE IS ONLY WEAKLY PRINTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
235597 SYRINGE 50ML LL CLEAR 14GA 1-1/4IN SYRINGE FMF BECTON DICKINSON, S.A. 1904243

Patients

Seq Age Sex Outcome Treatment
1 Other