FDA Adverse Event Malfunction Summary report: N

BD 60ML SYRINGE LUER-LOK TIP

MDR report key: 9764333 · Received February 27, 2020

Report

Report Number
1911916-2020-00192
Event Type
Malfunction
Date Received
February 27, 2020
Date of Event
February 4, 2020
Report Date
February 21, 2020
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
UDI-DI
30382903096535
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: 40 SAMPLES WERE RECEIVED. THEY CAME IN SEALED PACKAGING BLISTERS. THEY ARE 60 ML AND WERE VISUALLY INSPECTED AND NO DEFECTS WERE OBSERVED. THEY WERE TESTED FOR LEAKAGE AND THEY ALL PASSED. THE BARREL SCALE MARKING AND PACKAGING TOP WEB ARE AS 60 ML. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED WITH ZERO DEFECTS FOUND. NO QUALITY NOTIFICATIONS WERE WRITTEN FOR THIS BATCH, NOR FOR THE ASSOCIATED ASSEMBLY BATCHES. INVESTIGATION CONCLUSION: THIS IS THE 1ST COMPLAINT FOR LOT # 9120824 FOR THIS TYPE OF DEFECT OR SYMPTOM. THERE WAS NO DOCUMENTATION FOR THIS TYPE OF DEFECT DURING THE ENTIRE PRODUCTION RUN OF THIS BATCH. ROOT CAUSE DESCRIPTION: UNDETERMINED. RATIONALE: CAPA NOT REQUIRED AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD 60ML SYRINGE LUER-LOK¿ TIP LEAKED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO.: 309653 BATCH NO.: 9120824. 1 OF 3 - OCCURRENCE DATE (B)(6) 2020. IT WAS REPORTED THE PLUNGER SEAL FAILED CAUSING FLUID TO LEAK FROM THE PLUNGER. VERBATIM: PLUNGER SEAL FAILED CAUSING FLUID TO LEAKING FROM PLUNGER. THIS OCCURRED ONCE DURING TRANSPORTATION AND TWICE WHILE DRAWING FLUID INTO THE SYRINGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
230622 BD 60ML SYRINGE LUER-LOK TIP PISTON SYRINGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 309653 9120824 30382903096535

Patients

Seq Age Sex Outcome Treatment
1 Other