FDA Adverse Event Injury Summary report: N

WRIGHT MEDICAL MICROPORT RENAISSANCE MODULAR COCR NECK

MDR report key: 9763811 · Received February 26, 2020

Report

Report Number
MW5093329
Event Type
Injury
Date Received
February 26, 2020
Date of Event
February 1, 2013
Report Date
February 24, 2020
Manufacturer
MICROPORT / WRIGHT MEDICAL
Product Code
LZO
Adverse Event
Yes
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
AK, US
Reporter Occupation
003

Narratives

Description of Event or Problem · 1

PT ID: (B)(6); ON (B)(6) 2013, SHE UNDERWENT A RIGHT TOTAL HIP ARTHROPLASTY WITH A SIZE 10 REDUCED PROFEMUR RENAISSANCE STEM, SIZE 46 MM DYNASTY POROUS-COATED SHELL, 6.5 X 15MM CANCELLOUS BONE SCREW, SIZE 28 MM +0 DEGREES INNER DIAMETER DYNASTY A CLASS POLINER, A 28MM SHORT CERAMIC FEMORAL HEAD, AND A SHORT VV COBALT CHROME MODULAR NECK. SINCE HER ARTHROPLASTY, SHE HAS DEVELOPED NEW FATIGUE, DIMINISHED RESILIENCE, HEARING LOSS, AND NEW AUTOIMMUNE PROBLEMS AND ALLERGIES. IN (B)(6) OF 2019, SHE NOTED NEW SYMPTOMS OF RIGHT HIP PAIN WITH HIP FLEXION AGAINST RESISTANCE AND PAIN WITH HIP ABDUCTION WHEN SHE SKIS. THE METAL SUPPRESSION MRI OF RIGHT HIP SHOWED SUBTLE ANTERIOR FOREIGN BODY GRANULOMA OR FLUID COLLECTION. ON (B)(6) 2019, URINE COBALT LEVEL WAS 0.5 MCG/L AND THE BLOOD COBALT LEVEL WAS LESS THAN 0.5 MCG/L. NEUROQ ANALYSIS OF HER FDG PET BRAIN SCAN SHOWED ABNORMAL HYPOMETABOLISM INVOLVING 11 BRAIN CLUSTER REGIONS IN A PATTERN COMPATIBLE WITH CHRONIC TOXIC ENCEPHALOPATHY. ON (B)(6) 2020, THE RIGHT HIP WAS REVISED TO MICROPORT VV TIAIV ALLOY MODULAR NECK 8 AND A NEW DELTA CERAMIC 28MM +0 HEAD. THE OLD STEM WAS SOUND AND WAS IN ABOUT 15 DEGREES OF ANTEVERSION, THERE WAS GROSS CORROSION AT STEM END OF THE MODULAR COCR NECK AND ALSO METAL TRANSFER INTO BORE CERAMIC HEAD. POSTERIOR CAPSULE WAS INTACT, ANTERIOR CAPSULE WAS THICKENED IN TYPICAL ADVERSE REACTION TO METAL DEBRIS, ABDUCTION TENDONS WERE INTACT. PATHOLOGY REPORT OF FROZEN SECTION OF RIGHT HIP CAPSULE SYNOVIUM NOTABLE FOR FIBROTIC SYNOVIUM WITH AN ALVAL SCORE OF 3 OUT OF 10. FDA SAFETY REPORT ID# (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
220637 WRIGHT MEDICAL MICROPORT RENAISSANCE MODULAR COCR NECK PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC / POLYMER LZO MICROPORT / WRIGHT MEDICAL
220638 SCREW, FIXATION, BONE PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC / POLYMER HWC UNK
220639 WRIGHT MEDICAL MICROPORT RENAISSANCE MODULAR COCR NECK PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC / POLYMER LZO MICROPORT / WRIGHT MEDICAL

Patients

Seq Age Sex Outcome Treatment
1 68 YR Other| R