FDA Adverse Event Malfunction Summary report: N

BONE PIN, 3.2MM X 140MM, STERILE 2 PACK

MDR report key: 9757994 · Received February 26, 2020

Report

Report Number
3005985723-2020-00106
Event Type
Malfunction
Date Received
February 26, 2020
Date of Event
January 30, 2020
Report Date
May 14, 2020
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
UDI-DI
07613327412376
PMA / PMN Number
K112507
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

REPORTED ISSUE- SURGEON BEGAN PLACING PIN IN HIP FOR ARRAY PLACEMENT (NON OPERATIVE HIP FOR DA APPROACH). SURGEONS STATED HE BARELY BROKE BIN WHEN THE TIP BROKE OFF. PATIENT WAS X-RAYED AND TIP OF PIN WAS REMOVED FROM THE BONE. PRODUCT INSPECTION - AS PER THE IMAGE PROVIDED IN THE COMMUNICATION LOG , THE BROKEN PIN CAN BE SEEN AND CONFIRMED. ALSO THE DEVICE HAS BEEN DISCARDED BY THE HOSPITAL. PRODUCT HISTORY REVIEW ¿ REVIEW OF THE DEVICE HISTORY RECORDS INDICATE (B)(4) WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK ON 03/05/2019. NO NON-CONFORMANCES WERE IDENTIFIED DURING INSPECTION. COMPLAINT HISTORY REVIEW - A REVIEW OF COMPLAINTS IN CATSWEB AND TRACKWISE RELATED TO P/N 213527, LOT 48100219 SHOWS NO ADDITIONAL COMPLAINT RELATED TO THE FAILURE IN THIS INVESTIGATION. CONCLUSION ¿ THE FAILURE IS CONFIRMED VIA VISUAL INSPECTION OF THE IMAGE PROVIDED IN THE COMMUNICATION LOG. NO ADDITIONAL INVESTIGATION OR SPECIFIC ACTIONS ARE REQUIRED. IF ADDITIONAL INFORMATION IS RECEIVED THEN THE COMPLAINT WILL BE REOPENED. H3 OTHER TEXT : DEVICE NOT RETURNED.

Description of Event or Problem · 0

SURGEON BEGAN PLACING PIN IN HIP FOR ARRAY PLACEMENT (NON OPERATIVE HIP FOR DA APPROACH). SURGEONS STATED HE BARELY BROKE BIN WHEN THE TIP BROKE OFF. PATIENT WAS X-RAYED AND TIP OF PIN WAS REMOVED FROM THE BONE. CASE TYPE: THA.

Additional Manufacturer Narrative · 1

AS PART OF THE NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

SURGEON BEGAN PLACING PIN IN HIP FOR ARRAY PLACEMENT (NON OPERATIVE HIP FOR DA APPROACH). SURGEONS STATED HE BARELY BROKE BIN WHEN THE TIP BROKE OFF. PATIENT WAS X-RAYED AND TIP OF PIN WAS REMOVED FROM THE BONE. CASE TYPE: THA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
218022 BONE PIN, 3.2MM X 140MM, STERILE 2 PACK ORTHOPEDIC STEREOTAXIC INSTRUMENT OLO MAKO SURGICAL CORP. 213527 48100219 07613327412376

Patients

Seq Age Sex Outcome Treatment
1 31 YR Other