FDA Adverse Event
Other
Summary report: N
PRISMA CONTROL UNIT
MDR report key: 975719
·
Received January 4, 2007
Report
- Report Number
- 2087532-2007-00008
- Event Type
- Other
- Date Received
- January 4, 2007
- Date of Event
- December 6, 2006
- Report Date
- December 6, 2006
- Manufacturer
- GAMBRO DASCO
- Product Code
- FII
- Product Problem
- Yes
- Report Source
- Distributor report
- Reporter Location
- TN, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THERE WAS A PROBLEM WITH THE PRISMA MACHINE TAKING OFF TOO MUCH FLUID DURING A PT'S CONTINUOUS RENAL REPLACEMENT THERAPY (CRRT) SESSION. FACILITY'S INTENSIVE CARE NURSE HAD THE MACHINE PROGRAMMED TO REMOVE 210 ML/HOUR, HOWEVER FROM 1:00 AM TO 2:00 AM THE MACHINE INDICATED THAT THE FLUID REMOVAL WAS 406 ML. SHE PROGRAMMED THE MACHINE FOR ZERO FLUID REMOVAL, BUT NOTED IN THE NEXT HALF-HOUR THE MACHINE HAD ALREADY REMOVED 137 ML. THUS, THERE WAS AN EXCESSIVE FLUID REMOVAL OF 333 ML OVER A 1.5 HOUR TIME FRAME DURING A PT'S CRRT SESSION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRISMA CONTROL UNIT | INTENSIVE CARE HEMODIALYSIS | FII | GAMBRO DASCO | PRISMA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Other |