ENDOPATH ENDOSCOPIC MULTIFEED STAPLER
Report
- Report Number
- 1527736-1997-01188
- Event Type
- Malfunction
- Date Received
- June 10, 1997
- Date of Event
- May 12, 1997
- Report Date
- June 9, 1997
- Manufacturer
- ETHICON ENDO-SURGERY, INC.
- Product Code
- GAG
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
VISUAL INSPECTIONS & RESULTS: HEAD ROTATES, A,B, YES. NOSE SHROUD CRACKED/BROKEN, A,B, NO. STAPLES IN NOSE, A,B, YES. STAPLES IN THE TRACK, A,B, YES. TRIGGER ENGAGED WITH PRECOCK, A,B, YES. FUNCTIONAL TESTS & RESULTS: CYCLED INSTRUMENT, A,B, YES. ANALYSIS CONCLUSION: A,B)BASED UPON THE INQUIRY INFO RECEIVED, THE VISUAL EXAMINATION AND THE FUNCTIONAL TEST, IT WAS CONCLUDED THAT THE INSTRUMENTS WERE CONFORMIIGWITH REGARD TO STAPLES FEEDIND, FIRING AND FORMING. THE INSTRUMENTS WERE EXAMINED, WERE FOUND TO BE FUNCTIONAL, AND WERE CYCLED, FIRED, AND PROPERLY FORMED; A)THE REMAINING 17 STAPLES WITHOUT INCIDENT. B)THE REMAINING 12 STAPLES WITHOUT INCIDENT. NO CONCLUSION COULD BE REACHED AS TO WHY THE REPORTED INSTRUMENTS "DID NOT FIRE" AND "JAMMED" DURING SURGERY. A,B)EACH INSTRUMENT IS EVALUATED DURING ASSEMBLY PROCESS TO ENSURE THAT STAPLES FEED, FIRE AND FORM CORRECTLY. CO STRIVE TO UNDERSTAND EACH INCIDENT AS IT OCCURS TO CONTINUOUSLY IMPROVE CO'S PRODUCT'S.
DURING A T.E.P. PROCEDURE THE EMS USED TO SECURE MESH ON THE THIRD AND FOURTH FIRING DID NOT STAPLE. THE FIFTH TEST FIRING INTO A TOWEL DID NOT FIRE. THE SIX FIRING JAMMED AND A SECOND EMS WAS OPENED. DURING USE OF THE SECOND EMS, THE FOURTH AND FIFTH FIRING DID NOT STAPLE. THE SIXTH TEST FIRING INTO GAUZE DID FIRE AS DID THE SEVENTH AND EIGHTH. THE PROCEDURE WAS COMPLETED WITH THE SECOND EMS INSTRUMENT. THERE WAS NO CONSEQUENCE TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH ENDOSCOPIC MULTIFEED STAPLER | ENDOSCOPIC STAPLER | GAG | ETHICON ENDO-SURGERY, INC. | NA | H40K59 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR | Other |