FDA Adverse Event Malfunction Summary report: N

ENDOPATH ENDOSCOPIC MULTIFEED STAPLER

MDR report key: 97527 · Received June 10, 1997

Report

Report Number
1527736-1997-01188
Event Type
Malfunction
Date Received
June 10, 1997
Date of Event
May 12, 1997
Report Date
June 9, 1997
Manufacturer
ETHICON ENDO-SURGERY, INC.
Product Code
GAG
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

VISUAL INSPECTIONS & RESULTS: HEAD ROTATES, A,B, YES. NOSE SHROUD CRACKED/BROKEN, A,B, NO. STAPLES IN NOSE, A,B, YES. STAPLES IN THE TRACK, A,B, YES. TRIGGER ENGAGED WITH PRECOCK, A,B, YES. FUNCTIONAL TESTS & RESULTS: CYCLED INSTRUMENT, A,B, YES. ANALYSIS CONCLUSION: A,B)BASED UPON THE INQUIRY INFO RECEIVED, THE VISUAL EXAMINATION AND THE FUNCTIONAL TEST, IT WAS CONCLUDED THAT THE INSTRUMENTS WERE CONFORMIIGWITH REGARD TO STAPLES FEEDIND, FIRING AND FORMING. THE INSTRUMENTS WERE EXAMINED, WERE FOUND TO BE FUNCTIONAL, AND WERE CYCLED, FIRED, AND PROPERLY FORMED; A)THE REMAINING 17 STAPLES WITHOUT INCIDENT. B)THE REMAINING 12 STAPLES WITHOUT INCIDENT. NO CONCLUSION COULD BE REACHED AS TO WHY THE REPORTED INSTRUMENTS "DID NOT FIRE" AND "JAMMED" DURING SURGERY. A,B)EACH INSTRUMENT IS EVALUATED DURING ASSEMBLY PROCESS TO ENSURE THAT STAPLES FEED, FIRE AND FORM CORRECTLY. CO STRIVE TO UNDERSTAND EACH INCIDENT AS IT OCCURS TO CONTINUOUSLY IMPROVE CO'S PRODUCT'S.

Description of Event or Problem · 1

DURING A T.E.P. PROCEDURE THE EMS USED TO SECURE MESH ON THE THIRD AND FOURTH FIRING DID NOT STAPLE. THE FIFTH TEST FIRING INTO A TOWEL DID NOT FIRE. THE SIX FIRING JAMMED AND A SECOND EMS WAS OPENED. DURING USE OF THE SECOND EMS, THE FOURTH AND FIFTH FIRING DID NOT STAPLE. THE SIXTH TEST FIRING INTO GAUZE DID FIRE AS DID THE SEVENTH AND EIGHTH. THE PROCEDURE WAS COMPLETED WITH THE SECOND EMS INSTRUMENT. THERE WAS NO CONSEQUENCE TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH ENDOSCOPIC MULTIFEED STAPLER ENDOSCOPIC STAPLER GAG ETHICON ENDO-SURGERY, INC. NA H40K59

Patients

Seq Age Sex Outcome Treatment
1 41 YR Other