FDA Adverse Event Malfunction Summary report: N

INTELEPACS

MDR report key: 9751805 · Received February 25, 2020

Report

Report Number
9615916-2020-00002
Event Type
Malfunction
Date Received
February 25, 2020
Date of Event
February 7, 2020
Report Date
February 20, 2020
Manufacturer
INTELERAD MEDICAL SYSTEMS
Product Code
LLZ
UDI-DI
B228INTELEPACS0
PMA / PMN Number
K150707
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

N.A.

Description of Event or Problem · 1

A USER REPORTED THAT APPROXIMATELY 1000 STUDIES WERE LOST FROM THE PACS STORAGE AND A REFERRING PHYSICIAN WAS UNABLE TO VIEW IMAGES. RADIOLOGIST REPORTS FOR THESE DELETED STUDIES REMAINED AVAILABLE. PRELIMINARY ANALYSIS DETERMINED THE PROBLEM IS DUE TO A CONFIGURATION KEY THAT WAS ACTIVATED IN COMBINATION WITH A SOFTWARE BUG. WHEN THE FEATURE FORCEGLOBALSTUDYDELETION IS ENABLED, ANY SERIES DELETIONS THAT OCCUR ON A SERVER ARE AUTOMATICALLY ELEVATED TO A SYSTEM-LEVEL STUDY DELETION. THIS SYSTEM-LEVEL DELETION BYPASSES EXISTING MITIGATION POLICIES (I.E., ARCHIVE POLICIES) ON OTHER SERVERS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
212147 INTELEPACS PICTURE ARCHIVING AND COMMUNICATION SYSTEM LLZ INTELERAD MEDICAL SYSTEMS INTELEPACS 4-9-1 AND UP B228INTELEPACS0

Patients

Seq Age Sex Outcome Treatment
1