FDA Adverse Event
Malfunction
Summary report: N
INTELEPACS
MDR report key: 9751805
·
Received February 25, 2020
Report
- Report Number
- 9615916-2020-00002
- Event Type
- Malfunction
- Date Received
- February 25, 2020
- Date of Event
- February 7, 2020
- Report Date
- February 20, 2020
- Manufacturer
- INTELERAD MEDICAL SYSTEMS
- Product Code
- LLZ
- UDI-DI
- B228INTELEPACS0
- PMA / PMN Number
- K150707
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
N.A.
Description of Event or Problem · 1
A USER REPORTED THAT APPROXIMATELY 1000 STUDIES WERE LOST FROM THE PACS STORAGE AND A REFERRING PHYSICIAN WAS UNABLE TO VIEW IMAGES. RADIOLOGIST REPORTS FOR THESE DELETED STUDIES REMAINED AVAILABLE. PRELIMINARY ANALYSIS DETERMINED THE PROBLEM IS DUE TO A CONFIGURATION KEY THAT WAS ACTIVATED IN COMBINATION WITH A SOFTWARE BUG. WHEN THE FEATURE FORCEGLOBALSTUDYDELETION IS ENABLED, ANY SERIES DELETIONS THAT OCCUR ON A SERVER ARE AUTOMATICALLY ELEVATED TO A SYSTEM-LEVEL STUDY DELETION. THIS SYSTEM-LEVEL DELETION BYPASSES EXISTING MITIGATION POLICIES (I.E., ARCHIVE POLICIES) ON OTHER SERVERS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 212147 | INTELEPACS | PICTURE ARCHIVING AND COMMUNICATION SYSTEM | LLZ | INTELERAD MEDICAL SYSTEMS | INTELEPACS 4-9-1 AND UP | B228INTELEPACS0 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |