FDA Adverse Event Malfunction Summary report: N

BD 5ML SYRINGE LUER-LOK TIP

MDR report key: 9751406 · Received February 25, 2020

Report

Report Number
1213809-2020-00136
Event Type
Malfunction
Date Received
February 25, 2020
Date of Event
February 4, 2020
Report Date
March 19, 2020
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096498
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: ONE 5ML SYRINGE IN A OPENED BLISTER PACK FROM BATCH 7122534 (P/N 309646) WAS RECEIVED AND EVALUATED. IT WAS OBSERVED THE STOPPER WAS JAMMED BETWEEN THE PLUNGER ROD AND THE BARREL WALL, WHICH WAS REJECTABLE PER PRODUCT SPECIFICATION. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. POTENTIAL ROOT CAUSE FOR THE JAMMED STOPPER DEFECT IS ASSOCIATED WITH THE ASSEMBLY PROCESS. NO CORRECTIVE ACTIONS ARE NECESSARY BASED ON THE DEFECTIVE RATE IDENTIFIED. BATCH 7122534 IS CONSIDERED IN COMPLIANCE WITH OUR PRODUCT SPECIFICATION REQUIREMENTS.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO USE THE STOPPER WAS DISCOVERED TO BE DAMAGED WITH A BD 5ML SYRINGE LUER-LOK¿ TIP. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTED THAT THE STOPPER IS DAMAGED.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO USE THE STOPPER WAS DISCOVERED TO BE DAMAGED WITH A BD 5ML SYRINGE LUER-LOK¿ TIP. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTED THAT THE STOPPER IS DAMAGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
216139 BD 5ML SYRINGE LUER-LOK TIP PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 309649 7122534 30382903096498

Patients

Seq Age Sex Outcome Treatment
1 Other