FDA Adverse Event Other Summary report: N

*

MDR report key: 974955 · Received May 25, 2007

Report

Report Number
9610617-2007-00014
Event Type
Other
Date Received
May 25, 2007
Manufacturer
*
Product Code
EZO
Report Source
Manufacturer report

Narratives

Additional Manufacturer Narrative · 1

THE BLADE AND ATTACHED SHAFT (2 CM IN LENGTH TOTAL) WAS RETURNED FOR EVAL. THIS IS THE BROKEN PIECE THAT WAS RETRIEVED FROM THE PT. THE BLADE OF THE SHAFT SHOWS A LOT OF STRIATIONS ALONG THE FACE OF THE BLADE; IT APPEARS TO HAVE BEEN IN HEAVY USE. THERE APPEARS TO BE GOUGED METAL ON THE TIP AND THERE ARE SOME NICKS IN THE METAL ALONG THE EDGE OF THE CUTTING/BLADE SURFACE. THERE IS NO DEFECT OR CORROSION VISIBLE IN THE SHAFT METAL AT POINT OF BREAKAGE. THERE IS NO DEFECT TREND REPORTED ON THIS DEVICE. WE BELIEVE THE CONDITION OF THIS INSTRUMENT IS CONSISTENT WITH DAMAGE CAUSED BY MECHANICAL OVERLOAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 * * EZO * * *

Patients

Seq Age Sex Outcome Treatment
1