AEROSET ANALYZER
Report
- Report Number
- 1628664-2008-00006
- Event Type
- Malfunction
- Date Received
- January 8, 2008
- Date of Event
- December 13, 2007
- Report Date
- December 13, 2007
- Manufacturer
- ABBOTT MANUFACTURING, INC.
- Product Code
- JJE
- PMA / PMN Number
- K980367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
THE ABBOTT FIELD SERVICE REPRESENTATIVE (FSR) WAS DISPATCHED TO INSPECT THE INSTRUMENT. THE CUVETTE WASHER HEAD WAS ADJUSTED, THE DRYER TIP WAS REPLACED, ALL WASHER NOZZLES WERE STYLETTED, AND THE ICT MODULE PROBE WAS ADJUSTED. THE SAMPLE ARM SHAFT A AND B TUBING, ICT JOINT TUBING, AND ICT VALVE TO FITTING TUBING WERE REPLACED AS A PRECAUTION. THE TEXT STATES PROPER ANALYZER OPERATION VERIFICATIONS WERE PERFORMED AND PASSED. THE CUSTOMER RAN CONTROLS FOR MULTIPLE ASSAYS, WITH ALL CONTROLS WITHIN ACCEPTABLE RANGE. THE ISSUE IS ADDRESSED IN THE CLINICAL CHEMISTRY (CC) GLUCOSE REAGENT AND PACKAGE INSERTS UNDER SECTIONS: REAGENT HANDLING AND STORAGE AND SPECIMEN COLLECTION AND HANDLING. THE AEROSET SYSTEM OPERATIONS MANUAL (P/N 200154-101 -- NOVEMBER 2004) ALSO ADDRESSES THIS ISSUE IN THE SECTION ENTITLED: SECTION 7, OPERATIONAL PRECAUTIONS AND LIMITATIONS - RESULTS INTERPRETATION AND REPORTING. A MALFUNCTION OF THE SYSTEM OCCURRED TO CAUSE THE GENERATION OF A DISCREPANT GLUCOSE RESULT. THE EXACT CAUSE OF THE MALFUNCTION COULD NOT BE DETERMINED AS MULTIPLE COMPONENTS WERE REPLACED IN TROUBLESHOOTING THE ISSUE. THIS IS A FINAL REPORT.
(B)(4). DATE RECEIVED BY MANUFACTURER CHANGED FROM (B)(4) 2004 TO (B)(4) 2007; TYPOGRAPHICAL ERROR. DEVICE MANUFACTURE DATE CHANGED FROM 01/2000 TO 11/01/2000; TYPOGRAPHICAL ERROR. THIS IS A FINAL REPORT.
THE CUSTOMER STATES THAT MULTIPLE ASSAYS ARE GENERATING ABERRANT RESULTS ON THE AEROSET ANALYZER. ONE PT GENERATED AN INITIAL GLUCOSE ASSAY RESULT OF 33 MG/DL THAT RETESTED AT 129 AND 129 MG/DL. THERE IS NO IMPACT TO PT MANAGEMENT REPORTED. A SERVICE CALL WAS INITIATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AEROSET ANALYZER | AUTOMATED CHEMISTRY ANALYZER | JJE | ABBOTT MANUFACTURING, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK YR | LIST # 7D66-20 OR 3L82-20, LOT # UNK| CLINICAL CHEMISTRY GLUCOSE ASSAY| CLIN CHEM GLUCOSE LN:7D66 OR 3L82 LOT:NI |