JMS SYCLOC A.V. FISTULA NEEDLE SET
Report
- Report Number
- 3002807350-2007-00007
- Event Type
- Other
- Date Received
- January 2, 2008
- Date of Event
- November 13, 2007
- Report Date
- November 17, 2007
- Manufacturer
- JMS SINGAPORE PTE. LTD.
- Product Code
- FIE
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
AVF JOINT DETACHED FROM TUBE DURING DISCONNECTION OF AVF NEEDLE FROM FMC BLOODLINE AT THE END OF DIALYSIS PROCESS. END-USER DID NOT SEE ANY AIR POCKET AT BONDING AREA BEFORE USAGE. NO AIR EMBOLISM DURING TREATMENT. USUALLY, END-USER HOLDS AVF FEMALE CONNECTOR/JOINT AND TWIST THE PT MALE CONNECTOR FOR CONNECTION AND DISCONNECTION. REVIEW OF PRESERVED SAMPLES: PERFORMED VISUAL INSPECTION ON APPEARANCE OF PRESERVED SAMPLES (N=5EA). NO AIR POCKET, WHITE PATCHES, INSUFFICIENT BONDING OR POOR INSERTION WAS OBSERVED. PERFORMED JOINT TO TUBE PULL STRENGTH TEST ON PRESERVED SAMPLES (N=2EA). AVF JOINTS TO TUBE PULL STRENGTH WERE 8.8 KGF & 9.4 KGF. (JMSS QC SPECIFICATION JOINT + TUBE PULL TEST SPECIFICATION: >6 KGF.) REVIEW OF DEVICE HISTORY RECORDS (DHR): THE RECORDS WERE CHECKED AND NO DISCREPANCY WAS FOUND. NO PROCESS DEVIATION AND REWORK WAS ASSOCIATED TO THIS PRODUCT LOT. POSSIBLE ROOT CAUSE: SUCH COMPLAINT REPORTED CASE (AVF JOINT DETACHED FROM TUBE) MIGHT RELATE TO INSUFFICIENT COATING OF ADHESIVE ONTO TUBE-AVF JOINT BONDING AREA. BASED ON THE PRODUCTION TRACEABILITY, ALL THE TUBE-AVF JOINTS WERE BONDED BY AUTO-BONDING PROCESS ON HB MACHINE #4. MISALIGNMENT OF ADHESIVE COATING PIN MIGHT LEAD TO UNEVEN COATING OF ADHESIVE ONTO COATING CHUCK, AS A RESULT, INSUFFICIENT COATING MIGHT OCCUR AND CAUSING POOR INSERTION OF AVF JOINT TO TUBE. PRODUCTION INSPECTOR MIGHT HAVE OVERLOOKED THE DEFECT DURING 100% INSPECTION. CORRECTIVE ACTIONS: BY THE END OF DEC '07, PRODUCTION WILL CONVERT THE AUTO-BONDING PROCESS ON HB MACHINE #4 TO MANUAL BONDING PROCESS WHICH IS USING VALIDATED COATING JIG. MANUAL COATING IS MORE EFFICIENT IN TERMS OF COATING CONSISTENCY AND COATING LENGTH. ON 26-NOV-07, BRIEFED ALL RELATED OPERATORS AND INSPECTORS TO INCREASE THEIR AWARENESS FOR THE DEFECT DURING ASSEMBLY PROCESS AND INSPECTION.
LUER LOCK CAME OFF AT THE END OF TREATMENT; NO BLOOD LOSS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | JMS SYCLOC A.V. FISTULA NEEDLE SET | A.V. FISTULA NEEDLE | FIE | JMS SINGAPORE PTE. LTD. | 824-1501 | 070713531 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |