FDA Adverse Event Malfunction Summary report: N

FISCHER CONE BIOP EX MED

MDR report key: 9734374 · Received February 20, 2020

Report

Report Number
1216677-2020-00055
Event Type
Malfunction
Date Received
February 20, 2020
Date of Event
February 10, 2020
Report Date
February 1, 2023
Manufacturer
COOPERSURGICAL, INC.
Product Code
HGI
UDI-DI
00888937003710
PMA / PMN Number
K061651
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: INITIATED MANUFACTURER'S INVESTIGATION: NO SAMPLE RETURNED. REVIEW DHR. DISTRIBUTION HISTORY THE 15006-02 FISHER CONE ASSEMBLY WAS PURCHASED FROM GEOTEC, INC. AS AN OEM FINISHED PRODUCT AND ISSUED TO WORK ORDER(B)(4) AS CSI PART NUMBER 900-151 AND COMPLETED 4/23/2019. MANUFACTURING RECORD REVIEW: THE DHR FOR THIS UNIT WAS REVIEWED AND NO NON-CONFORMITIES, RELATED TO THE COMPLAINT CONDITION, WERE NOTED. INCOMING INSPECTION REVIEW: A REVIEW OF THE INCOMING INSPECTION RECORD COULD NOT BE PERFORMED AS THE RECORD COULD NOT BE LOCATED AT THE TIME OF THIS INVESTIGATION. IF THE INCOMING INSPECTION RECORD SHOULD BE LOCATED GOING FORWARD, IT WILL BE REVIEWED, AND THIS COMPLAINT AMENDED ACCORDINGLY. SERVICE HISTORY RECORD: SERVICE HISTORY RECORD NOT APPLICABLE TO THIS PRODUCT. HISTORICAL COMPLAINT REVIEW: A REVIEW OF THE PRODUCT TWO-YEAR HISTORY INDICATED DID NOT REVEALED SIMILAR REPORTED COMPLAINT CONDITIONS. PRODUCT RECEIPT: THE COMPLAINT UNIT WAS NOT RETURNED TO COOPERSURGICAL. VISUAL EVALUATION: EVALUATION OF THE COMPLAINT UNIT COULD NOT BE COMPLETED AT THE COMPLAINT UNIT HAS NOT BEEN RETURNED TO COOPERSURGICAL. IF THE UNIT SHOULD BE RETURNED AT A LATER DATE, IT WILL BE EVALUATED, AND ANY FINDINGS WILL BE APPENDING TO THIS INVESTIGATION. FUNCTIONAL EVALUATION: EVALUATION OF THE COMPLAINT UNIT COULD NOT BE COMPLETED AT THE COMPLAINT UNIT HAS NOT BEEN RETURNED TO COOPERSURGICAL. IF THE UNIT SHOULD BE RETURNED AT A LATER DATE, IT WILL BE EVALUATED, AND ANY FINDINGS WILL BE APPENDING TO THIS INVESTIGATION. ROOT CAUSE: DEFINITIVE ROOT CAUSE IS INDETERMINABLE HOWEVER, PREVIOUS TESTING PERFORMED IN 2011 IN TRYING TO REPLICATE REPORTED EVENTS OF THE WIRE "BURNING" OR "SNAPPING" INDICATED THE PRODUCT PERFORMED AS INTENDED, THE TESTING WAS REPEATED IN MARCH OF 2019 AND RESULTED IN THE SAME MANNER. CORRECTIVE ACTIONS: COOPERSURGICAL WILL CONTINUE TO MONITOR THIS COMPLAINT CONDITION FOR TRENDS. NO FURTHER CORRECTIVE ACTION IS NECESSARY, AS THE COMPLAINT CONDITION WAS NOT CONFIRMED. NO FURTHER TRAINING REQUIRED AT THIS TIME. WAS THE COMPLAINT CONFIRMED? NO.

Description of Event or Problem · 0

CUSTOMER STATED: "DURING EXCISION BIOPSY, WIRE BROKE AND CHARRED (BLUE) INSULATION ON EXCISOR". 1216677-2020-00055-1 900-151 FISCHER CONE BIOP EX MED (B)(4).

Additional Manufacturer Narrative · 1

COOPERSURGICAL INC. IS CURRENTLY INVESTIGATING THE REPORTED COMPLAINT CONDITION. REF (B)(4).

Description of Event or Problem · 1

DETAILS SURROUNDING EVENT- DURING EXCISION BIOPSY, WIRE BROKE AND CHARRED (BLUE) INSULATION ON EXCISOR. SURGEON REPORTS A TOTAL OF THREE INCIDENTS WHERE WIRE HAS BROKEN DURING PROCEDURE. THIS IS FIRST REPORTED INCIDENT. REF (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
199649 FISCHER CONE BIOP EX MED FISCHER CONE BIOP EX MED HGI COOPERSURGICAL, INC. 900-151 264280 00888937003710

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other