FDA Adverse Event Malfunction Summary report: N

SILKAM BLACK 0 (3.5) 75CM HS30

MDR report key: 9727523 · Received February 19, 2020

Report

Report Number
3003639970-2020-00084
Event Type
Malfunction
Date Received
February 19, 2020
Date of Event
January 27, 2020
Report Date
February 24, 2020
Manufacturer
B. BRAUN SURGICAL SA
Product Code
GAP
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

ANALYSIS AND RESULTS: THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE-BATCH. WE MANUFACTURED AND DISTRIBUTED IN THE MARKET (B)(4) UNITS OF THIS CODE-BATCH. THERE ARE NO UNITS IN OUR STOCK. WE HAVE RECEIVED (B)(4) CLOSED SAMPLES FOR ANALYSIS. WE HAVE TESTED THE KNOT PULL TENSILE STRENGTH OF THE SAMPLES RECEIVED AND THE RESULTS FULFIL THE REQUIREMENTS OF THE EUROPEAN PHARMACOPOEIA (EP): 2.54 KGF IN AVERAGE AND 2.42 KGF IN MINIMUM (EP REQUIREMENTS: 2.24 KGF IN AVERAGE AND 1.33 KGF IN MINIMUM) REVIEWED THE BATCH MANUFACTURING RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFIL USP/EP AND B.BRAUN SURGICAL REQUIREMENTS. REMARKS: WHEN WORKING WITH SILKAM SUTURE MATERIALS GREAT CARE SHOULD BE TAKEN TO ENSURE THAT THE USE OF SURGICAL INSTRUMENTS, SUCH AS TWEEZERS AND NEEDLE HOLDERS DO NOT DAMAGE THE MATERIAL BY BEING PINCHED OR KINKED. FINAL CONCLUSION: ALTHOUGH THE RESULTS OF THE SAMPLES RECEIVED FULFIL THE SPECIFICATIONS OF EUROPEAN PHARMACOPOEIA/ B. BRAUN SURGICAL SPECIFICATIONS, WE TAKE NOTE OF THIS INCIDENCE IN ORDER TO ASSESS IF NEW OR ADDITIONAL ACTIONS ARE NEEDED. ACTIONS ON PRODUCT DISTRIBUTED OF THIS REFERENCE/BATCH: BASED ON THE CONCLUSION DERIVED FROM INVESTIGATION, IT IS NOT REQUIRED TO MAKE ACTIONS IN DISTRIBUTED PRODUCT. CORRECTIVE/PREVENTIVE ACTIONS: ACCORDING TO OUR INTERNAL PROCEDURES, THERE IS NO NEED TO ESTABLISH CORRECTIVE OR PREVENTIVE ACTIONS. NEVERTHELESS, THIS COMPLAINT IS RECORDED FOR TRENDING ANALYSIS TO ASSESS IF ACTIONS ARE NEEDED IN THE FUTURE.

Additional Manufacturer Narrative · 1

PMA/510K: REPORTED DEVICE NOT MARKETED IN THE U.S., HOWEVER, SIMILAR DEVICES OR DEVICES THAT SHARE COMPONENTS, RAW MATERIALS, PROCESS METHODS OR OTHER TECHNOLOGICAL CHARACTERISTICS ARE REGISTERED WITHIN THE U.S. K990089. WHEN ADDITIONAL INFORMATION/INVESTIGATION RESULTS BECOME AVAILABLE, RESULTS WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS AN ISSUE WITH SILKAM BLACK. THE CUSTOMER REPORTED THAT THE THREAD BROKE DURING SURGERY. THERE IS NO PATIENT INFORMATION AVAILABLE, BUT HAS BEEN REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
191735 SILKAM BLACK 0 (3.5) 75CM HS30 OTHER SUTURE GAP B. BRAUN SURGICAL SA C0764620 619153

Patients

Seq Age Sex Outcome Treatment
1