FDA Adverse Event Injury Summary report: N

STEM: AMISTEM H HA COATED LAT STEM SIZE 7

MDR report key: 9726276 · Received February 19, 2020

Report

Report Number
3005180920-2020-00069
Event Type
Injury
Date Received
February 19, 2020
Date of Event
January 20, 2020
Report Date
February 19, 2020
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030804236
PMA / PMN Number
K093944
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED BY MEDACTA REGULATORY AFFAIRS DEPARTMENT ON 18 FEBRUARY 2020: LOT 111711: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 18-JUL-2011. EXPIRATION DATE: 2016-06-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS MANAGER: FEMORAL COMPONENT (STEM AND HEAD) REVISION PERFORMED 8 YEARS AFTER PRIMARY CEMENTLESS TOTAL HIP ARTHROPLASTY IN AN ELDERLY ((B)(6) YEAR OLD) MAN. NO INFORMATION CONCERNING PATIENT GENERAL HEALTH STATUS AND THE PRESENCE OF COMORBIDITIES IS AVAILABLE. IN THE RADIOGRAPHIC IMAGE PROVIDED, REGIONS OF REDUCED BONE DENSITY AROUND THE PROXIMAL PROFILE OF THE STEM AND SIGNS OF STRESS SHIELDING ARE VISIBLE. ASEPTIC LOOSENING IS A POSSIBLE LITERATURE DESCRIBED LONG-TERM ADVERSE EVENT AFTER PRIMARY CEMENTLESS HIP ARTHROPLASTIES AND CAUSES ARE OFTEN UNKNOWN. THE REASON OF THIS FAILURE CANNOT BE DETERMINED. PRELIMINARY INVESTIGATION PERFORMED BY R&D PROJECT MANAGER: PRELIMINARY VISUAL INVESTIGATION HIGHLIGHTS RESIDUAL HA ON PROXIMAL FEMORAL STEM AND SIGNS OF OSSEOINTEGRATION ON DISTAL STEM.

Description of Event or Problem · 1

REVISION SURGERY DUE TO STEM LOOSENING, STEM AND HEAD REVISED. PRIMARY WAS PERFORMED IN (B)(6) 2012. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
193747 STEM: AMISTEM H HA COATED LAT STEM SIZE 7 CEMENTLESS HIP STEM LZO MEDACTA INTERNATIONAL SA 01.18.147 111711 07630030804236

Patients

Seq Age Sex Outcome Treatment
1 81 YR Required Intervention