HUDSON AQUAPAK 728 SW,760 ML W/028 ADAPTOR
Report
- Report Number
- 1417411-2020-00006
- Event Type
- Malfunction
- Date Received
- February 18, 2020
- Date of Event
- January 22, 2020
- Report Date
- January 31, 2020
- Product Code
- OGG
- PMA / PMN Number
- K141214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
QN#(B)(4). CORRECTED DATA: SECTION D.1.- BRAND NAME CORRECTED TO HUDSON AQUAPAK 728 SW,760 ML W/028 ADAPTOR SECTION D.2.- COMMON DEVICE NAME CORRECTED TO HUMIDIFIER NEBULIZER KIT SECTION D.2B.- PROCODE CORRECTED TO OGG SECTION D.4.- CATALOG# CORRECTED TO 037-28 ONE EMPTY 037-00, 760 ML, WATER BOTTLE LOT 19F110 WAS RECEIVED FOR EVALUATION. NO ADAPTOR WAS RECEIVED FOR EVALUATION. THE FRONT NEBULIZER OUTLET PORT WAS INTACT. THE TOP ADAPTOR PORT ON THE WATER BOTTLE HAD BEEN PUNCTURED CLEANLY WITHOUT PORT INVERSION. THE NUMBER "4" WAS EMBOSSED IN THE PLASTIC OF THE BOTTOM OF THE WATER BOTTLE. NO VISUAL DEFECTS OBSERVED. A DEVICE HISTORY RECORD REVIEW WAS PERFORMED AND NO RELEVANT FINDINGS WERE IDENTIFIED. THE PRODUCT INSTRUCTIONS FOR USE STATE, "THREAD THE NEBULIZER ADAPTOR ONTO THE RESERVOIR AND TIGHTEN." AS NO ADAPTOR SAMPLE IS AVAILABLE FROM THE CUSTOMER, THE COMPLAINT COULD NOT BE CONFIRMED AND THE PROBABLE CAUSE COULD NOT BE DETERMINED FROM THE AVAILABLE INFORMATION. TELEFLEX WILL CONTINUE TO MONITOR AND TREND FOR REPORTS OF THIS NATURE.
CUSTOMER REPORTED THAT STAFF WERE UNABLE TO OR HAD DIFFICULTY PUNCTURING THE TOP OF THE BOTTLE. NO PATIENT HARM REPORTED.
(B)(4).
CUSTOMER REPORTED THAT STAFF WERE UNABLE TO OR HAD DIFFICULTY PUNCTURING THE TOP OF THE BOTTLE. NO PATIENT HARM REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 182506 | HUDSON AQUAPAK 728 SW,760 ML W/028 ADAPTOR | HUMIDIFIER NEBULIZER KIT | OGG | 19F110 | |||
| 182507 | HUDSON AQUAPAK 728 SW,760 ML W/028 ADAPTOR | HUMIDIFIER NEBULIZER KIT | OGG | 19F110 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NONE REPORTED.| NONE REPORTED. |