FDA Adverse Event Malfunction Summary report: N

HUDSON AQUAPAK 728 SW,760 ML W/028 ADAPTOR

MDR report key: 9723163 · Received February 18, 2020

Report

Report Number
1417411-2020-00009
Event Type
Malfunction
Date Received
February 18, 2020
Date of Event
January 22, 2020
Report Date
January 31, 2020
Product Code
OGG
PMA / PMN Number
K141214
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

QN# (B)(4). CORRECTED DATA: SECTION D.1.- BRAND NAME CORRECTED TO HUDSON AQUAPAK 728 SW,760 ML W/028 ADAPTOR. SECTION D.2.- COMMON DEVICE NAME CORRECTED TO HUMIDIFIER NEBULIZER KIT. SECTION D.2B.- PROCODE CORRECTED TO OGG. SECTION D.4.- CATALOG# CORRECTED TO 037-28. ONE EMPTY 037-00, 760 ML, WATER BOTTLE LOT 19F110 WAS RECEIVED FOR EVALUATION. NO ADAPTOR WAS RECEIVED FOR EVALUATION. THE WATER BOTTLE ARRIVED WITH BOTH PORTS PUNCTURED. THE FRONT NEBULIZER OUTLET PORT HAD BEEN PUNCTURED CLEANLY WITHOUT PORT INVERSION. THE TOP ADAPTOR PORT ON THE WATER BOTTLE WAS INVERTED SUCH THAT THE PORT HAD NOT RUPTURED CLEANLY. THE NUMBER "4" WAS EMBOSSED IN THE PLASTIC OF THE BOTTOM OF THE WATER BOTTLE. HANDWRITING ON THE SIDE OF THE BOTTLE APPEARED TO BE IN BLACK MARKER INK AND READ, "PUNCTURED BY HAND". NO OTHER VISUAL DEFECTS OBSERVED. A DEVICE HISTORY RECORD REVIEW WAS PERFORMED AND NO RELEVANT FINDINGS WERE IDENTIFIED. THE PRODUCT INSTRUCTIONS FOR USE STATE, "THREAD THE NEBULIZER ADAPTOR ONTO THE RESERVOIR AND TIGHTEN." AS NO ADAPTOR SAMPLE IS AVAILABLE FROM THE CUSTOMER, THE COMPLAINT COULD NOT BE CONFIRMED AND THE PROBABLE CAUSE COULD NOT BE DETERMINED FROM THE AVAILABLE INFORMATION. TELEFLEX WILL CONTINUE TO MONITOR AND TREND FOR REPORTS OF THIS NATURE.

Description of Event or Problem · 0

CUSTOMER REPORTED THAT STAFF WERE UNABLE TO OR HAD DIFFICULTY PUNCTURING THE TOP OF THE BOTTLE. NO PATIENT HARM REPORTED.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED THAT STAFF WERE UNABLE TO OR HAD DIFFICULTY PUNCTURING THE TOP OF THE BOTTLE. NO PATIENT HARM REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
182201 HUDSON AQUAPAK 728 SW,760 ML W/028 ADAPTOR HUMIDIFIER NEBULIZER KIT OGG 19F110
182202 HUDSON AQUAPAK 728 SW,760 ML W/028 ADAPTOR HUMIDIFIER NEBULIZER KIT OGG 19F110

Patients

Seq Age Sex Outcome Treatment
1 NONE REPORTED.| NONE REPORTED.