FDA Adverse Event Injury Summary report: N

3.5MM 90 DEG DFOS PLATE, 4MM OFFSET, 4-HOLE

MDR report key: 9717158 · Received February 17, 2020

Report

Report Number
3006460162-2020-00029
Event Type
Injury
Date Received
February 17, 2020
Date of Event
August 14, 2019
Report Date
February 13, 2020
Manufacturer
ORTHOPEDIATRICS, CORP
Product Code
HRS
PMA / PMN Number
K083286
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

REFERENCE: (B)(4). COMPLAINT SAMPLE WAS NOT RETURNED FOR EVALUATION, HOSPITAL SCRAPPED IT. REPORTED EVENT WAS NOT CONFIRMED AFTER EVALUATING THE CUSTOMER SUBMITTED RADIOGRAPH. DHR REVIEW WAS UNABLE TO BE PERFORMED AS THE LOT NUMBER OF THE INVOLVED DEVICE IS UNKNOWN. THERE WERE NO RECENT DESIGN CHANGES PRIOR TO MANUFACTURING. RISK MANAGEMENT FILE REVIEW WAS DEEMED APPROPRIATE. REVIEW OF THE COMPLAINT HISTORY DETERMINED THAT NO FURTHER ACTION IS REQUIRED AS NO TRENDS WERE IDENTIFIED. ROOT CAUSE WAS UNABLE TO BE DETERMINED, THE PLATE DID NOT APPEAR TO BE BENT OR FRACTURED IN THE CUSTOMER SUBMITTED RADIOGRAPH. CONCOMITANT DEVICES: 3.5MM LOCKING CORTICAL SCREW W/ T15 HEX, SIZE 22MM: ITEM NUMBER; 00-0903-2622, LOT NUMBER; UNKNOWN. 3.5MM LOCKING CORTICAL SCREW W/ T15 HEX, SIZE 24MM: ITEM NUMBER; 00-0903-2624, LOT NUMBER; UNKNOWN. 3.5MM LOCKING CORTICAL SCREW W/ T15 HEX, SIZE 50MM: ITEM NUMBER; 00-0903-2650, LOT NUMBER; UNKNOWN. 3.5MM SELF TAPPING CORTICAL SCREW W/ T15 HEX, SIZE 22MM: ITEM NUMBER; 00-0903-2522, LOT NUMBER; UNKNOWN. 3.5MM SELF TAPPING CORTICAL SCREW W/ T15 HEX, SIZE 44MM: ITEM NUMBER; 00-0903-2644, LOT NUMBER; UNKNOWN. 3.5MM SELF TAPPING CORTICAL SCREW W/ T15 HEX, SIZE 40MM: ITEM NUMBER; 00-0903-2640, LOT NUMBER; UNKNOWN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DISTAL FEMORAL OSTEOTOMY SYSTEM PLATE ROTATED IN THE DISTAL PORTION OF THE BONE. A REVISION SURGERY TOOK PLACE APPROXIMATELY SIX WEEKS AFTER IMPLANTATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
181370 3.5MM 90 DEG DFOS PLATE, 4MM OFFSET, 4-HOLE DISTAL FEMORAL OSTEOTOMY SYSTEM HRS ORTHOPEDIATRICS, CORP

Patients

Seq Age Sex Outcome Treatment
1 14 YR Required Intervention