FDA Adverse Event Malfunction Summary report: N

BD SYRINGE LUER-LOK TIP

MDR report key: 9716770 · Received February 17, 2020

Report

Report Number
1213809-2020-00096
Event Type
Malfunction
Date Received
February 17, 2020
Date of Event
January 26, 2020
Report Date
March 9, 2020
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096283
PMA / PMN Number
K941562
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY: ONE PHOTO WAS RECEIVED AND EVALUATED. THE PHOTO DEPICTED A LOOSE 1ML LL SYRINGE. THE BARREL WAS OBSERVED TO HAVE SEVERAL SMALL PARTICLES POTENTIALLY EMBEDDED IN THE WALL OPPOSITE 0.4, 0.5 AND POSSIBLY 0.7ML MARKINGS. DUE TO THE PHOTO BEING OUT OF FOCUS IT WAS NOT POSSIBLE TO IDENTIFY THE FOREIGN MATTER PARTICLES, HOW MANY THERE WERE, NOR THEIR SIZE. A PHYSICAL SAMPLE OR ADDITIONAL PHOTOS ARE REQUIRED TO IDENTIFY THE FM. POTENTIAL ROOT CAUSE FOR THE EMBEDDED FOREIGN MATTER DEFECT IS ASSOCIATED WITH THE MOLDING PROCESS. NO CORRECTIVE ACTIONS ARE NECESSARY BASED ON THE DEFECTIVE RATE IDENTIFIED. BATCH 8211583 IS CONSIDERED IN COMPLIANCE WITH OUR PRODUCT SPECIFICATION REQUIREMENTS. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT RED PARTICLES WERE FOUND EMBEDDED IN THE BD SYRINGE LUER-LOK¿ TIP PLASTIC DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WE RECEIVED A COMPLAINT ABOUT RED PARTICLES IN A 1 ML SYRINGE USED BY US AND PREPARED A TRANSPARENT COLOR PREPARATION." "THE PARTICLES APPEAR TO BE EMBEDDED IN THE PLASTIC THAT KIND OF MADE THE SYRINGE."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT RED PARTICLES WERE FOUND EMBEDDED IN THE BD SYRINGE LUER-LOK¿ TIP PLASTIC DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WE RECEIVED A COMPLAINT ABOUT RED PARTICLES IN A 1 ML SYRINGE USED BY US AND PREPARED A TRANSPARENT COLOR PREPARATION." "THE PARTICLES APPEAR TO BE EMBEDDED IN THE PLASTIC THAT KIND OF MADE THE SYRINGE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
181956 BD SYRINGE LUER-LOK TIP PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 309628 8211583 30382903096283

Patients

Seq Age Sex Outcome Treatment
1 Other