FDA Adverse Event Malfunction Summary report: N

THERMOCOOL SMART TOUCH SF BI-DIRECTIONAL NAVIGATION CATHETER

MDR report key: 9712701 · Received February 14, 2020

Report

Report Number
2029046-2020-00244
Event Type
Malfunction
Date Received
February 14, 2020
Date of Event
January 22, 2020
Report Date
January 22, 2020
Manufacturer
BIOSENSE WEBSTER INC
Product Code
LPB
UDI-DI
10846835010183
PMA / PMN Number
P030031
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE BIOSENSE WEBSTER, INC. PRODUCT ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION AND NOTED ON FEBRUARY 15, 2020 THAT THERE WAS RED-BROWNISH MATERIAL ON THE DOME. SINCE THE EVENT DESCRIPTION REPORTED THAT THROMBUS WAS FOUND ADHERED TO THE TIP OF THE CATHETER WHEN IT WAS REMOVED FROM THE PATIENT¿S BODY, THE REDDISH-BROWN MATERIAL OBSERVED BY THE BIOSENSE WEBSTER, INC. PRODUCT ANALYSIS LAB WAS ASSESSED AS THROMBUS AND REMAINS REPORTABLE. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETED AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. MANUFACTURER'S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE THE CAVOTRICUSPID ISHMUS (CTI) WAS ABLATED, AND THROMBUS WAS FOUND ADHERED TO THE TIP OF THE THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER (LOT# 30280027M) WHEN IT WAS REMOVED FROM THE SHEATH. THE CATHETER WAS REPLACED WITH A NEW THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER (LOT # 30276828M) AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THE DEVICE WAS VISUALLY INSPECTED AND A RED-BROWNISH MATERIAL WAS OBSERVED ON THE DOME. A CLOSER SECOND INSPECTION WAS PERFORMED AND THROMBUS WAS IDENTIFIED AS THE MATERIAL ON THE DOME. AN ELECTRICAL TEST WAS PERFORMED ON THE CATHETER AND IT WAS FOUND WITHIN SPECIFICATIONS. NO ELECTRICAL MALFUNCTION WAS OBSERVED. ADDITIONALLY, THE CATHETER WAS TESTED ON THE GENERATOR AND THE TEMPERATURE AND IMPEDANCE VALUES WERE OBSERVED WITHIN SPECIFICATIONS. THEN, A COOL FLOW PUMP TEST WAS PERFORMED, AND IT WAS FOUND WITHIN SPECIFICATIONS. THE CATHETER WAS IRRIGATING CORRECTLY. NO IRRIGATION ISSUES WERE OBSERVED. A MANUFACTURING RECORD EVALUATION WAS PERFORMED AND NO INTERNAL ACTION RELATED TO THE COMPLAINT WAS FOUND DURING THE REVIEW. THE CUSTOMER COMPLAINT WAS CONFIRMED. THE ROOT CAUSE OF THE THROMBUS REPORTED BY THE CUSTOMER COULD BE RELATED TO THE USAGE OF THE DEVICE DURING THE PROCEDURE; HOWEVER, THIS CANNOT BE CONCLUSIVELY DETERMINED. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. MANUFACTURER'S REFERENCE NUMBER: PC-(B)(4).

Additional Manufacturer Narrative · 0

MANUFACTURER'S REFERENCE NUMBER: (B)(4) INITIALLY THIS EVENT WAS ASSESSED AS MDR REPORTABLE FOR A THROMBUS ISSUE. DURING REVIEW ON(B)(6) 2021, A CORRECTION WAS NOTED TO THE ASSESSMENT AS THIS EVENT SHOULD HAVE BEEN ASSESSED AS CHAR WHICH IS NOT MDR REPORTABLE. CHAR IS A PHYSICAL PHENOMENON OF RADIO FREQUENCY ENERGY DELIVERY AND CAN BE THE NORMAL RESULT OF THE ABLATION PROCESS. THE PRESENCE OF CHAR ON THE ELECTRODES DOES NOT REPRESENT A SERIOUS INJURY IN ITSELF, NOR IS IT NECESSARILY THE RESULT OF DEVICE MALFUNCTION. THEREFORE, UPDATED UNDER ¿H6. MEDICAL DEVICE PROBLEM CODE¿ FROM ¿COAGULATION IN DEVICE OR DEVICE INGREDIENT¿ TO ¿DEVICE CONTAMINATION WITH BODY FLUID¿.

Additional Manufacturer Narrative · 1

INITIAL REPORTER PHONE: (B)(6). IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. BIOSENSE WEBSTER MANUFACTURER'S REFERENCE NUMBER (B)(4) HAS TWO COMPLAINTS THAT ARE RELATED TO THE SAME INCIDENT. MANUFACTURER'S REF. NO: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE THE CAVOTRICUSPID ISHMUS (CTI) WAS ABLATED, AND THROMBUS WAS FOUND ADHERED TO THE TIP OF THE THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER (LOT# 30280027M) WHEN IT WAS REMOVED FROM THE SHEATH. THE CATHETER WAS REPLACED WITH A NEW THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER (LOT # 30276828M) AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. AT THE END OF THE PROCEDURE, WHEN THE REPLACEMENT CATHETER WAS REMOVED FROM THE PATIENT¿S BODY, THROMBUS WAS FOUND ATTACHED TO THE TIP. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ERROR MESSAGES WERE OBSERVED ON ANY BIOSENSE WEBSTER, INC. EQUIPMENT. THE PHYSICIAN CONFIRMED THE PATIENT DID NOT EXHIBIT ANY NEUROLOGICAL SYMPTOM SINCE THE PROCEDURE WAS COMPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
175421 THERMOCOOL SMART TOUCH SF BI-DIRECTIONAL NAVIGATION CATHETER CARDIAC ABLATION PERCUTANEOUS CATHETER LPB BIOSENSE WEBSTER INC D134805 30280027M 10846835010183

Patients

Seq Age Sex Outcome Treatment
1 THMCL SMTCH SF BID, TC, D-F.| UNKNOWN BRAND SHEATH.| UNK_SMARTABLATE GENERATOR.