FDA Adverse Event Malfunction Summary report: N

CRANIAL DRILL BITS AND ACCESSORIES

MDR report key: 9711061 · Received February 14, 2020

Report

Report Number
2183456-2020-00004
Event Type
Malfunction
Date Received
February 14, 2020
Date of Event
June 5, 2017
Report Date
February 14, 2020
Manufacturer
AD-TECH MEDICAL INSTRUMENT CORP.
Product Code
HBE
Removal / Correction Number
2183456-08212019-R-004
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

MDR 2183456-2020-00004 HAS BEEN FILED PAST THE 30-DAY TIMEFRAME. THROUGH THE 2019 FDA INSPECTION, IT WAS IDENTIFIED THAT WHILE AD-TECH WAS EVALUATING REPORTABILITY FOR ACTUAL EVENTS IN WHICH DEATH OR SERIOUS INJURY HAD OCCURRED, AD-TECH HAD FAILED TO SUBMIT EVENTS IN WHICH A MALFUNCTION WAS REPORTED AND COULD RESULT IN A SERIOUS INJURY IF THE EVENT WERE TO RECUR. AS PART OF THE INVESTIGATION, A TWO (2) YEAR RETROSPECTIVE REVIEW OF COMPLAINTS WAS PERFORMED TO DETERMINE WHICH COMPLAINTS REQUIRED SUBMISSION OF AN MDR. MDR-2183456-2020-00004 WAS SUBMITTED AS A CONSERVATIVE MEASURE DUE TO THE RECALLS TIED TO THIS ISSUE.

Description of Event or Problem · 1

A USER FACILITY REPORTED THAT THE DRILL BIT SEIZED IN THE DRILL SLEEVE GUIDE, POSSIBLY DUE TO IMPROPER HANDLING BY THE USER FACILITY. BACK UP TOOLS WERE USED AND THE CASE PROCEEDED WITHOUT PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
173396 CRANIAL DRILL BITS AND ACCESSORIES DRILL SLEEVE GUIDE HBE AD-TECH MEDICAL INSTRUMENT CORP.

Patients

Seq Age Sex Outcome Treatment
1 Other