FDA Adverse Event Injury Summary report: N

MYSPINE MYSPINE MC DRILL BASED GUIDE L04

MDR report key: 9710937 · Received February 14, 2020

Report

Report Number
3005180920-2020-00059
Event Type
Injury
Date Received
February 14, 2020
Date of Event
January 16, 2020
Report Date
March 19, 2020
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
PQC
UDI-DI
07630040717533
PMA / PMN Number
K173472
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

VISUAL INSPECTION PERFORMED BY R&D PROJECT MANAGER: THE MYSPINE BONE MODELS (REF. 7.0704-7.0705) AND THE MYSPINE GUIDES (REF. 7.0724-7.0725) OF L4 AND L5 RESPECTIVELY HAVE BEEN INSPECTED FOR A DEEPER INVESTIGATION OF THE COMPLAINT. THE FITTING BETWEEN THE BONE MODELS AND THE GUIDES OF L4 AND L5 RESULTS ARE FULLY CONFORMING: THE GUIDE CAN BE DOCKED AND REMOVED FROM THE BONE MODEL WITHOUT ANY ISSUE (SEE IMAGE 1 AND 2 ATTACHED). FURTHERMORE, FROM THE VISUAL INSPECTION, IT CAN BE NOTICED THAT THERE IS NO IMPINGEMENT BETWEEN THE GUIDE AND THE UPPER FACET JOINTS. DUE TO THE FACT, THE FACETS OF THE UPPER VERTEBRAE (L3 FOR L4 GUIDE AND L4 FOR L5 GUIDE) NOT NEEDED TO BE RESECTED.

Additional Manufacturer Narrative · 1

PATIENT MATCH PLANNING REVIEW AS PER YOUR REQUEST, WE ANALYZED EACH STEP OF THE MYSPINE PROCESS. IMAGES QC: THE IMAGES WE RECEIVED WERE SUBJECTED TO QUALITY CONTROL, THE RESULT OF WHICH WAS POSITIVE. RECONSTRUCTION: THE RECONSTRUCTION IS PRECISE AND CORRECT, ALL AREAS WERE CLEARLY IDENTIFIABLE. THE RECONSTRUCTED PROFILES FOLLOW ACCURATELY THE CONTOUR OF THE VERTEBRAE: PLANNING: THE SURGEON VALIDATED OUR FIRST PLANNING PROPOSAL. THE GUIDES WERE DESIGNED ACCORDING TO THE TRAJECTORY OF THE SCREWS IN THE PLANNING VALIDATED BY THE SURGEON. CONCLUSIONS OUR ANALYSIS OF THE MYSPINE PROCESS OF THIS CASE FOUND NO DEVIATIONS FROM THE STANDARD PROCEDURES. EACH STEP HAS BEEN PERFORMED CORRECTLY. ADDITIONAL DEVICE INVOLVED: MYSPINE 7.0725 MYSPINE MC DRILL BASED GUIDE L05 (K173472) LOT. 02494S. ALL THE ITEMS INVOLVED ARE RELATED TO LOT NUMBER 02494S BATCH REVIEW PERFORMED ON 28 JANUARY 2019: LOT 02494S: (B)(4) GUIDES MANUFACTURED AND RELEASED ON 02-JAN-2020. NO ANOMALIES FOUND RELATED TO THE PROBLEM.

Description of Event or Problem · 1

THE SURGEON WAS NOT ABLE TO SET THE MYSPINE MC GUIDE CORRECTLY AT L4- L5. THE SURGEON FINISHED THE SURGERY WITHOUT USING THE MYSPINE MC GUIDE. DUE TO THIS EVENT THE SURGERY WAS PROLONGED ABOUT 1 HOUR. TOTAL SURGERY TIME ABOUT 200 MINS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
174262 MYSPINE MYSPINE MC DRILL BASED GUIDE L04 DRILL GUIDE FOR SPINE SURGERY PQC MEDACTA INTERNATIONAL SA 02494S 07630040717533

Patients

Seq Age Sex Outcome Treatment
1 93 YR Other