Narratives

Additional Manufacturer Narrative · 1

THE CEILING COVER CLAMP RING WAS BROKEN DUE TO EXCESSIVE TIGHTENING (TORQUE > 0.25 N/M) OF THE TWO ATTACHMENT SCREWS. PRIOR TO THE INCIDENT, THE LAST MAQUET PERFORMED MAINTENANCE RECORD ON THIS SURGICAL LIGHT DATES BACK TO 11/04. AS FOLLOW-UP, AN INTEGRITY CHECK OF EACH HLX 2000 CLAMP RING USED AT THE FACILITY WAS PERFORMED. REPLACEMENTS WERE MADE, WHEN NECESSARY.