BARDEX® ALL-SILICONE FOLEY CATHETER
Report
- Report Number
- 1018233-2020-00997
- Event Type
- Malfunction
- Date Received
- February 13, 2020
- Date of Event
- December 24, 2019
- Report Date
- March 24, 2020
- Manufacturer
- C.R. BARD, INC. (COVINGTON) -1018233
- Product Code
- EZL
- UDI-DI
- 00801741029745
- PMA / PMN Number
- K040504
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NURSE
Narratives
THE REPORTED EVENT WAS CONFIRMED. THE DEVICE DID NOT MEET SPECIFICATIONS. THE PRODUCT WAS INTENDED TO BE USED FOR TREATMENT PURPOSES. THE PRODUCT WAS INFLUENCED BY THE REPORTED FAILURE. VISUAL EVALUATION OF THE RETURNED SAMPLE NOTED ONE OPENED (WITHOUT ORIGINAL PACKAGING), SILICONE FOLEY CATHETER. VISUAL INSPECTION OF THE SAMPLE NOTED NO OBVIOUS VISIBLE DEFECTS. THE CATHETER BALLOON WAS ATTEMPTED TO BE INFLATED WITH 10 ML METHYLENE BLUE SOLUTION (3 DROPS 1% AQ METHYLENE BLUE PER 100ML DISTILLED WATER), AND IMMEDIATELY LEAKED FROM A PINHOLE (MEASURING 0.012 SQ MM) OVER THE INFLATION LUMEN AT THE BIFURCATION POINT. THIS IS OUT OF SPECIFICATION PER INSPECTION PROCEDURE, WHICH STATES, "CUTS IN LUMENS ARE NOT ALLOWED." ALTHOUGH THE REPORTED EVENT WAS CONFIRMED, THE ROOT CAUSE COULD NOT BE DETERMINED. A POTENTIAL ROOT CAUSE FOR THIS FAILURE COULD BE, ¿INSERTS WITH EDGES (OPERATOR HITS THE TUBE AGAINST THE BOTTOM INSERT WHEN REMOVING THE PIECE FROM THE MOLD).¿ THE DEVICE HISTORY RECORD WAS REVIEWED AND FOUND NOTHING THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THE INSTRUCTIONS FOR USE WERE FOUND ADEQUATE AND STATE THE FOLLOWING: ¿PEEL TO OPEN C. R. BARD, INC. COVINGTON, GA 30014 1-800-526-4455 WWW.BARDMEDICAL.COM BARDEX® ALL-SILICONE FOLEY CATHETER CATHETER SHAFT MADE ENTIRELY OF SILICONE ELASTOMER PD-50205 11/00 CAUTION: FEDERAL (U.S.A.) LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN. STERILE: UNLESS PACKAGE IS OPENED OR DAMAGED. WARNING: DO NOT USE OINTMENTS OR LUBRICANTS HAVING A PETROLATUM BASE. THEY WILL DAMAGE SILICONE AND MAY CAUSE BALLOON TO BURST. CAUTION: DO NOT ASPIRATE URINE THROUGH DRAINAGE FUNNEL WALL. STORAGE: STORE CATHETERS AT ROOM TEMPERATURE AWAY FROM DIRECT EXPOSURE TO LIGHT, PREFERABLY IN THE ORIGINAL BOX. SINGLE USE ONLY. DO NOT REUSE. DO NOT RESTERILIZE. FOR UROLOGICAL USE ONLY. VALVE TYPE: USE LUER SLIP SYRINGE. DO NOT USE NEEDLE. RECOMMENDED INFLATION CAPACITIES 3CC BALLOON: USE 5ML STERILE WATER 5CC BALLOON: USE 10ML STERILE WATER 30CC BALLOON: USE 35ML STERILE WATER DO NOT EXCEED RECOMMENDED CAPACITIES. WARNING: AFTER USE, THIS PRODUCT MAY BE A POTENTIAL BIOHAZARD. HANDLE AND DISPOSE OF IN ACCORDANCE WITH APPLICABLE LOCAL, STATE, AND FEDERAL LAWS AND REGULATIONS. NOTE: AGGRESSIVE TRACTION, PARTICULARLY IN THE PRESENCE OF SUTURING, IS NOT RECOMMENDED FOR 100% SILICONE FOLEY CATHETERS. BARD AND BARDEX ARE REGISTERED TRADEMARKS OF C. R. BARD, INC. OR AN AFFILIATE. TO DEFLATE CATHETER BALLOON: GENTLY INSERT A SYRINGE IN THE CATHETER VALVE. NEVER USE MORE FORCE THAN IS REQUIRED TO MAKE THE SYRINGE ¿STICK¿ IN THE VALVE. IF YOU NOTICE SLOW OR NO DEFLATION, RE-SEAT THE SYRINGE GENTLY. ALLOW THE BALLOON TO DEFLATE SLOWLY ON ITS OWN. DO NOT ASPIRATE OR MANUALLY ACCELERATE THE DEFLATION OF THE BALLOON. IF PERMITTED BY HOSPITAL PROTOCOL, THE VALVE ARM MAY BE SEVERED. IF THIS FAILS, CONTACT ADEQUATELY TRAINED PROFESSIONAL FOR ASSISTANCE, AS DIRECTED BY HOSPITAL PROTOCOL. SHOULD BALLOON RUPTURE OCCUR, CARE SHOULD BE TAKEN TO ASSURE THAT ALL BALLOON FRAGMENTS HAVE BEEN REMOVED FROM THE PATIENT. VISUALLY INSPECT THE PRODUCT FOR ANY IMPERFECTIONS OR SURFACE DETERIORATION PRIOR TO USE." H11:SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD.
THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE, A SUPPLEMENTAL REPORT WILL BE FILED. H11:SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD. H3 OTHER TEXT : THE DEVICE WAS NOT RETURNED.
IT WAS REPORTED THAT THE CATHETER FELL OUT OF THE PATIENT ON THE NEXT DAY OF USE.
IT WAS REPORTED THAT THE CATHETER FELL OUT OF THE PATIENT ON THE NEXT DAY OF USE.
THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE, A SUPPLEMENTAL REPORT WILL BE FILED. THE DEVICE WAS NOT RETURNED.
IT WAS REPORTED THAT THE CATHETER FELL OUT OF THE PATIENT ON THE NEXT DAY OF USE.