FDA Adverse Event Malfunction Summary report: N

CRANIAL DRILL BITS AND ACCESSORIES

MDR report key: 9705921 · Received February 13, 2020

Report

Report Number
2183456-2019-00026
Event Type
Malfunction
Date Received
February 13, 2020
Date of Event
April 2, 2018
Report Date
February 13, 2020
Manufacturer
AD-TECH MEDICAL INSTRUMENT CORP.
Product Code
HBE
UDI-DI
00841823106520
PMA / PMN Number
K170442
Removal / Correction Number
2183456-08212019-R-004
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

MDR 2183456-2019-00026 HAS BEEN FILED PAST THE 30-DAY TIMEFRAME. THROUGH THE 2019 FDA INSPECTION, IT WAS IDENTIFIED THAT WHILE AD-TECH WAS EVALUATING REPORTABILITY FOR ACTUAL EVENTS IN WHICH DEATH OR SERIOUS INJURY HAD OCCURRED, AD-TECH HAD FAILED TO SUBMIT EVENTS IN WHICH A MALFUNCTION WAS REPORTED AND COULD RESULT IN A SERIOUS INJURY IF THE EVENT WERE TO RECUR. AS PART OF THE INVESTIGATION, A TWO (2) YEAR RETROSPECTIVE REVIEW OF COMPLAINTS WAS PERFORMED TO DETERMINE WHICH COMPLAINTS REQUIRED SUBMISSION OF AN MDR. MDR-2183456-2019-00026 WAS SUBMITTED AS A CONSERVATIVE MEASURE DUE TO THE RECALLS TIED TO THIS ISSUE.

Description of Event or Problem · 1

AN AD-TECH CLINICAL SPECIALIST ATTENDING A CASE AT A USER FACILITY REPORTED THAT THE DRILL BIT SEIZED IN THE DRILL SLEEVE GUIDE, POSSIBLY DUE TO OVERTIGHTENING THE DRILL STOP, BENDING THE BIT. THE SECOND BIT WAS USED AND THE CASE PROCEEDED WITHOUT PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172398 CRANIAL DRILL BITS AND ACCESSORIES DRILL SLEEVE GUIDE HBE AD-TECH MEDICAL INSTRUMENT CORP. DSG-6.3-090-2.4N 712121 00841823106520

Patients

Seq Age Sex Outcome Treatment
1 Other