NEUROSIGN
Report
- Report Number
- 3011755356-2020-00002
- Event Type
- Malfunction
- Date Received
- February 13, 2020
- Date of Event
- January 14, 2020
- Report Date
- January 17, 2020
- Manufacturer
- THE MAGSTIM COMPANY LIMITED
- Product Code
- ETN
- PMA / PMN Number
- K923056
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE DEVICE WAS RETURNED TO MAGSTIM FOR INVESTIGATION. DURING THE INVESTIGATION, WHEN INCREASING (CLOCKWISE) THE CURRENT POTENTIOMETER ON THE N100 TO 5MA (4.9), THE STIMULATOR PULSE WAS SEEN TO FLUCTUATE. REDUCING (ANTI-CLOCKWISE) THE CURRENT POTENTIOMETER VERY SLIGHTLY CAUSED A BIG DECREASE IN THE AMPLITUDE OF THE STIMULATOR PULSE. IT IS A CHARACTERISTIC OF THE STIMULATOR PULSE IN A N100 TO FLUCTUATE SLIGHTLY, HOWEVER WHEN THE DECREASE IN AMPLITUDE IS TO THE AMOUNT OBSERVED, THIS INDICATES THAT THE CURRENT POTENTIOMETER IS FAULTY. SHOULD THE SURGEON BE USING THE N100 AT MAXIMUM CURRENT (5MA), IT IS LIKELY THAT THEY EXPERIENCED THIS FLUCTUATION IN STIMULATOR PULSE AMPLITUDE, CAUSING THE INTERMITTENT RESPONSE DURING STIMULATION. PROBABLE ROOT CAUSE - FAULTY CURRENT POTENTIOMETER (P/N475048), WHILST THE DEVICE HAD EXCEEDED ITS LIFETIME.
MAGSTIM'S PRODUCT/CLINICAL SPECIALIST CONTACTED THE HOSPITAL ON THE ISSUE. FEEDBACK RECEIVED FROM THE THEATRE MANAGER SUGGESTED THAT USER ERROR CAUSED THE ISSUE. IT APPEARS THAT WHEN THE DEVICE IS USED WHEN A REPRESENTATIVE OF MAGSTIM IS PRESENT THAT NO ISSUES ARE REPORTED. MAGSTIM'S PRODUCT/CLINICAL SPECIALIST ADVISED THE THEATRE MANAGER THAT, BECAUSE OF THE FACT THAT THE MACHINE DID NOT ALARM WHEN STIMULATED, AND THE POST-OPERATIVE ASSESSMENT CONFIRMED NO INJURY, THE MACHINE PROBABLY WORKED AS INTENDED I.E. IF IT WAS THE NERVE THAT WAS STIMULATED, THE MACHINE WOULD HAVE ALARMED. ALTHOUGH MAGSTIM'S PRODUCT/CLINICAL SPECIALIST CONTACTED THE HOSPITAL FOR ADVICE, THE HOSPITAL DID NOT WANT A VISIT BY MAGSTIM FOR TRAINING OR ADVICE AND INSTEAD DECIDED TO DISCARD THE PRODUCT'S USE. THE MAGSTIM COMPANY LTD. IS, AS PER YOUR REQUEST, CURRENTLY IN THE PROCESS OF SETTING UP A WEBTRADER ACCOUNT FOR MAGSTIM INC. USA. DURING THIS TRANSITION TIME, PLEASE ACCEPT THIS SUBMISSION AS OF THE MAGSTIM COMPANY LTD. AND MAGSTIM INC. USA BOTH. EXEMPTION NUMBER E2018009. (B)(4).
THYROID SURGERY. SURGEON ADVISED THAT WHEN STIMULATING THE RESPONSE WAS INTERMITTENT. SURGEON ALSO ADVISED THAT HE WAS STIMULATING THE RECURRENT LARYNGEAL NERVE (RLN) AND THERE WAS NO RESPONSE PROVIDED BY THE NERVE MONITOR. SURGEON THEN ALARMED THE THEATRE MANAGER THAT HE HAD CUT THE RLN AND APPEARED WORRIED ABOUT PERMANENT DAMAGE. POST-SURGERY ASSESSMENT ON THE PATIENT CONFIRMED THAT THERE WAS NO DAMAGE CAUSED TO THE NERVE AND PATIENT WAS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 171266 | NEUROSIGN | NEUROSIGN 100 | ETN | THE MAGSTIM COMPANY LIMITED | 100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | BIPOLAR PROBE 3601-00| LANTERN LARYNGEAL ELECTRODE (LLE) 6/7 MM |