TPRLC XR MP T1 PPS 16X117MM
Report
- Report Number
- 0001825034-2020-00686
- Event Type
- Malfunction
- Date Received
- February 12, 2020
- Date of Event
- January 29, 2020
- Report Date
- June 16, 2020
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- OQG
- PMA / PMN Number
- K120030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- 003
Narratives
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS CONFIRMED. UPON VISUAL INSPECTION IT WAS FOUND THE STERILE BARRIER HAD BEEN BROKEN. FURTHER INSPECTION OF THE RETURNED DEVICE REVEALED THERE WAS DEBRIS INSIDE THE STERILE PACKAGING THAT IS CONSISTENT WITH THE APPEARANCE OF FOAM DEBRIS FROM THE FOAM PACKAGING INSIDE THE STERILE BARRIER AND BLACK DEBRIS CONSISTENT WITH POROUS MATERIAL FROM THE DEVICE. DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. THESE PRODUCTS WERE LIKELY CONFORMING WHEN THEY LEFT ZIMMER BIOMET CONTROL. THE ROOT CAUSE OF THE REPORTED ISSUE IS ATTRIBUTED TO TRANSIT DAMAGE CAUSING THE FOAM PACKAGING AND POROUS MATERIAL FROM DEVICE TO BECOME ABRADED AND SHED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
(B)(4). CONCOMITANT MEDICAL PRODUCTS: TPRLC 133 T1 PPS HO 16X152MM CAT# 51-104160 LOT# 6047712, TPRLC 133 T1 PPS HO 9X137MM CAT# 51-104090 LOT# 6057369, TPRLC XR MP T1 PPS 14X113MM CAT# 51-145140 LOT# 3545229, TPRLC XR MP T1 PPS 14X113MM CAT# 51-145140 LOT# 2884150, TPRLC XR MP T1 PPS 17X119MM CAT# 51-145170 LOT# 3730608, TPRLC 133 MP TYPE1 PPS SO 17.0 CAT# 51-106170 LOT# 3564680, TPRLC 133 MP TYPE1 PPS SO 17.0 CAT# 51-106170 LOT# 3669984. FOREIGN COUNTRY: (B)(6). MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034-2020-00679, 0001825034-2020-00680, 0001825034-2020-00681, 0001825034-2020-00682, 0001825034-2020-00683, 0001825034-2020-00684, 0001825034-2020-00685.
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
IT WAS REPORTED THAT DEBRIS WAS FOUND IN STERILE PACKAGING WHILE INVESTIGATING STOCK. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 165071 | TPRLC XR MP T1 PPS 16X117MM | PROSTHESIS, HIP | OQG | ZIMMER BIOMET, INC. | NI | 3029799 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |