PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2953144-2007-01263
- Event Type
- Injury
- Date Received
- December 19, 2007
- Report Date
- November 25, 2007
- Manufacturer
- ABBOTT VASCULAR - VASCULAR SOLUTIONS
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICE WAS NOT RETURNED TO ABBOTT VASCULAR; THEREFORE, WE WERE NOT ABLE TO PERFORM A DEVICE ANALYSIS. THE LOT NUMBER WAS NOT PROVIDED; THEREFORE, A DEVICE HISTORY RECORD REVIEW COULD NOT BE PERFORMED. LITERATURE SEARCH ARTICLE: NEW ENGLAND SOCIETY FOR VASCULAR SURGERY 2007; 45: 1095-1101. WRITTEN BY W. ANTHONY LEE, MD, TITLED "TOTAL PERCUTANEOUS ACCESS FOR ENDOVASCULAR AORTIC ANEURYSM REPAIR ("PRECLOSE" TECHNIQUE)".
DEVICE MALFUNCTION: UNK. TIME OF SYMPTOMS/AE: DURING VESSEL CLOSURE. SYMPTOMS/AE: PERSISTENT BLEEDING REQUIRING SURGICAL REPAIR. THIS WAS NOTED THROUGH A PERIODIC ARTICLE REVIEW THAT ASSESSED TOTAL PERCUTANEOUS ACCESS FOR ENDOVASCULAR AORTIC ANEURYSM REPAIR. A PRECLOSE TECHNIQUE WAS USED INVOLVING DEPLOYMENT OF THE PROGLIDE DEVICE BEFORE PROCEDURAL SHEATH INSERTION. A TOTAL OF 559 PROGLIDE DEVICES WERE USED TO CLOSE 279 FEMORAL ARTERIES; THERE WERE FAILURES, MAINLY DUE TO OBESITY, DEVICE MALFUNCTION, SEVERE CALCIFICATION AND FAULTY ARTERIAL PUNCTURES. DURING THE VESSEL CLOSURE PROCEDURE OF THE COMMON FEMORAL ARTERY WITH THE PROGLIDE, PERSISTENT BLEEDING WAS NOTED REQUIRING SURGICAL CLOSURE. INFORMATION REGARDING THE CAUSE FOR THE FAILURE TO ACHIEVE HEMOSTASIS WAS NOT SPECIFIED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | MGB | ABBOTT VASCULAR - VASCULAR SOLUTIONS | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK YR | Required Intervention | SHEATH: 24F |