BD PHASEAL Y-SITE CONNECTOR C80
Report
- Report Number
- 3003152976-2020-00049
- Event Type
- Malfunction
- Date Received
- February 11, 2020
- Date of Event
- January 20, 2020
- Report Date
- March 26, 2020
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- ONB
- UDI-DI
- 30382905153045
- PMA / PMN Number
- K123213
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION: TWO PHOTOS WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION OF THE PHOTOS, NO LEAKAGE CAN BE OBSERVED. FOUR RETAINED SAMPLES WERE EVALUATED, IN ALL CASES THE PRODUCT FUNCTIONED AS INTENDED AND NO LEAKAGES WERE OBSERVED. PRODUCT UNDERGOES BOTH VISUAL AND FUNCTIONAL TESTING THROUGHOUT MANUFACTURING TO ENSURE THE QUALITY AND FUNCTIONALITY OF THE DEVICE. A DEVICE HISTORY REVIEW WAS PERFORMED FOR LOT 1905101, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO IDENTIFY A ROOT CAUSE RELATED TO OUR MANUFACTURING PROCESS AT THIS TIME.
IT WAS REPORTED THAT THE BD PHASEAL¿ Y-SITE CONNECTOR C80 LEAKED FROM THE NOZZLE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: "LEAKAGE AT THE RINCE NOZZLE, IT'S RECURRING".
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE BD PHASEAL¿ Y-SITE CONNECTOR C80 LEAKED FROM THE NOZZLE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: "LEAKAGE AT THE RINCE NOZZLE, IT'S RECURRING"
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 160511 | BD PHASEAL Y-SITE CONNECTOR C80 | INTRAVASCULAR ADMINISTRATION SET | ONB | BECTON DICKINSON, S.A. | 515304 | 1905101 | 30382905153045 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |